Pular para o conteúdo principal
GE Healthcare · CTLightSpeed

LightSpeed 4.X - GRE

Symptom

Customer reports a squealing noise coming from inside the console. See long description for more details. Customer reports a squealing noise coming from inside the console. If the noisy fan is inside the HP computer, DARC, IG, or SDDA, then the correct module will have to be replaced, as these fans are not a FRU.

Applies to

Models
LightSpeed
Product
LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2376909 FAN ASSY - DUAL 120MM FANS See long description Part #: 2376909 Part Type: GP CONSUMABLES Part Weight: 2.18 LBR Repairable: No Returnable: No Hazardous Material: N Description: FAN ASSY - DUAL 120MM FANS FRU: Yes Status: Orderable Comments: FLS - ADDED PER GRE RSPL 5.12.02 MS27898 FAN ASSY - DUAL 120MM FANS USED ON H16. MS13452 030520 RISK: FUNCTIONALITY FREQUENCY: MEDIUM PDR REQUESTED IM05 REQUESTED
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.