GE Healthcare · CTHiSpeed25382216016180001035331960012033968

"Scan hardware Stopped Scan" with DAS power errors.

Symptom

"Scan hardware Stopped Scan" with DAS power errors. "Scan hardware Stopped Scan" with DAS power errors

Error message

Error: 253822 HPDAS Power supply auxiliary channels out of tolerance See Long Description for complete error Error: 253822 HPDAS Power supply auxiliary channels out of tolerance Mon Dec 6 12:12:30 2004 Error: 253822 Suite: RHCT Host: FEP Process: CoordinateMode File: fepin.c 1.61 Method: FEPIN_dma_done Line: 1822 Function: Data Acquisition : Scan Data Acquisition Scan: 16016/2/5 Type: Helical Scan Seq: 180001 FEP DSP 1:HPDAS Power supply auxiliary channels out of tolerance View #2 Values in volts. 753: +5: 0.035331 <==== +6 Analog Supply Too Low 754: -5: 4.960012 755: +5: 5.033968 756: +15: 26.452396 <==== +15 Analog Supply Too High 757: -15: 15.327940 Note: Pay attention to the specific Power Supply values. Different values relate to different causes.

Applies to

Models
HiSpeed
Product
HiSpeed CT/i 4.X HiSpeed CT/i 4.x with Zeus HSA Gantry and SGI Indigo Console HiSpeed CT/i with:HSA GANTRY CT/I (RHAPSODE) CONSOLE (Indigo Computer) Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2302509-2 SIGNAL BRUSH BLOCK FOR HSA/CTI SYSTEMS See long description for details Part #: 2302509-2 Part Type: GP CONSUMABLES Part Weight: 0.62 LBR Repairable: No Returnable: No Hazardous Material: N Description: SIGNAL BRUSHBLOCK - REPLACEMENT CT/I FRU: Yes Status: Orderable Comments: FLS - 8-7-03, CONSUMABLE COLLECTION PER ANJANA ON PROMOTION THROUGH JUNE 30, 2004 DO NOT CHANGE PRICE PER SHAWN (MAYND) DO NOT DISCOUNT - PER SHAKUN Note: 2302509 FOR H1 (LIGHTSPEED) SYSTEMS That is, DO NOT mix and match ETC and Schliefring Signal Brush Blocks. Any matched pair of Signal Brush Blocks may be installed with either Power Brush Block. See HSA/CTI Non EMC and EMC Gantry Slipping Comparison in Direction 46-018316 for any updates to this information.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.