GE Healthcare · CTLightSpeed23000089550002230000069291443700050003
LightSpeed 4.X - GRE
Symptom
Scan abort during any scan with external x-ray interlocks not satisfied.
Error message
Error: 230000895 Scan Seq Id: #####View number sanity check failed: buffer number expected view number 1 actual view number 0 See Long Description for complete error: Error: 230000895 Scan Seq Id: XXXXXView number sanity check failed: buffer number expected view number 1 actual view number 0 Thu Apr 13 22:25:54 2006 Error: 230000895 Host: darc Process: dataacq File: dip_thread.cxx Line: 1422 Function: Data Acquisition : Scan Data Acquisition Scan Type: Cine Scan: 50002/1/48 Scan Seq Id: 319000View number sanity check failed: buffer number 0 expected view number 1 actual view number 0 Do Not Use - use one of the other Error: 230000069 Two, or more, unexpected events were detected during scan data transfer. The 230000069 Error has several varation see below. Sun Sep 5 11:10:31 2004 Error: 230000069 Host: darc Process: dataacq File: dip_thread.cxx Line: 396 Function: Data Acquisition : Scan Data Acquisition Scan Type: Scout Scan: 4167/1/1 Scan Seq Id: 1000 Two, or more, unexpected events were detected during scan data transfer. Bit Count Event description 3 1 TAXI modem violation 5 1 View length error Mon Dec 15 10:40:11 2008 span style="COLOR: #ff0000"> Error: 230000069 Host: darc P
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 5144058-500 Upgrade kit - GOC3 to GOC4. DARC2 is not included. Order proper DARC2 separately - REFER to SN11063012 See Long Description for Details. 5144058-500 Upgrade Kit GOC3 to GOC4 - without DARC2 Part #: 5144058-500 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: Service Kit - SDDA and GOC3 16 MDAS 16,8,4 GDAS DARC Replacement kit, JARRELL FRU: Yes Status: Orderable Comments: ** FE ASSIST ** This part number no longer contains a DARC2 it must be ordered in addition per Your system type. REFER to SN11063012a(July 26 2011 212001261)
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.