Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT260005510230020202230020203200002352230020204

LightSpeed 7.x (VCT, VCT Select, Pro32)

Symptom

Scanner shuts down on its own during scanning. Scanner shuts down on its own during scanning

Error message

Error: 260005510 Task 'darc_init' terminated prematurely See long description for full error Sat Feb 2 07:54:18 2008 Error: 260005510 Host: ct01 Process: darc_init File: procDARC.cxx Line: 84 Task 'darc_init' terminated prematurely

Error: 230020202 DARC applications cannot start. Terminating. See long description for details Error: 230020202 DARC applications cannot start. Terminating. Tue Aug 25 16:15:46 2009 Error: 230020202 Host: darc Process: darc_init File: darc_init.cxx Line: 54 DARC applications cannot start. Terminating.

Error:230020203 DARC applications were not shut down from a previous startup. This is most likely due to an abnormal shutdown and/or reboot on the OC computer See Long description for error Tue Jul 21 22:50:49 2009 Error:230020203 Host: darc Process: darc init File: darc init.cxx Line: 148 DARC applications were not shut down from a previous startup. This is most likely due to an abnormal shutdown and/or reboot on the OC computer.

Error: 200002352 A HOST_SHUTDOWN was initiatied by the unexpected termination of darc_init. See long description for full error Sat Feb 2 07:54:18 2008 Error: 200002352 Host: ct01 Process: cupMonitor File: ssProcStop.c Line: 76

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues