GE Healthcare · CT182614182607
Lightspeed 1.X - Octane
Symptom
Table will not move up or down during a elevation table reference. Table will not move up or down during a elevation table reference
Error message
Error: 182614 Tilt drive is not stopping within 1500 msec. Expected stopping position XX (.01 deg) Current position YY (.01 deg). See Long Description for error details and example Host: ETC Process: Tilt File: tmcShutDown.c 1.31 1/29/93 Line: 121 Function: PATIENT ANATOMY POSITIONING : Elevation - Tilt Scan: 5478/4/0 Type: None/Unknown Exception Level: Pri/Soft Time: 16:25:56:115 Log Series: 151 HW/SW error: Tilt drive is NOT stopping within 1500 msec. Expected Stopping Position -50 (.01 deg). Current Position -23 (.01 deg). Tilt Direction: 0 (0 = BCK 1 = FWD). Error: 182607 Host: ETC Process: Tilt File: CmdMotion.c 1.21 2/17/95 Line: 139 Function: PATIENT ANATOMY POSITIONING : Elevation - Tilt Scan: 7839/2/0 Type: None/Unknown Exception Level: Pri/Soft Time: 12:12:01:160 Log Series: 617 HW/SW error: Velocity tracking error. Actual Velocity 0 (.01 deg/sec) Expected Velocity 100 (.01 deg/sec) Tol 80 (.01 deg/sec) Cur Pos -2181 (.01 deg) Tilt Dir: 1 (0 = BCK 1 = FWD). INFO: Probable Problem could be 180V Power is LOW.
Applies to
- Product
- Lightspeed 1.X - Octane 10.5_2.8.2I_H1.3M4 10.4.1_2.8.2I_H1.3M4 - obsolete Lightspeed 1.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config showprods|grep Apps note: if check_config does not work then the software is not up to date with all FMI's. Check FMI history and verify that all patches have been loaded since last software reload. Ref FMI 25269 List of Supported Console Types Octane with Pegasus Recon hardware. may have an 02 SBC - Run df -k to see if the sbc is partitioned. List of softwares the product can run 10.5_2.8.2I_H1.3M4 10.4.1_2.8.2I_H1.3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3176T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 46-264368G1 / T3202DD ETC Board. See Long Description for more details. ETC Board 46-264368G1 / T3202DD Part #: 46-264368G1 Part Type: GP REPAIRABLES Part Weight: 0.00 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: ETC. BOARD FRU: Yes Status: Orderable Comments: Not Available
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required