Pular para o conteúdo principal
GE Healthcare · CTLightSpeed2300190022600050031818850152230006001

LightSpeed 3.X (Ultra, Plus, QX/i) - Octane

Symptom

"Scan hardware Stopped Scan" with scan data errors. "Scan hardware Stopped Scan" with scan data errors

Error message

Error: 230019002 The scan disk I/O layer failed to consume a response from the scan disk server. See LD for complete error Tue Nov 2 03:19:30 2004 Error: 230019002 Host: sgredct1 Process:scanfilemgr File: sdio proxy.c Line:174 The scan disk I/O layer failed to consume a response from the scan disk server. function:consume sequence:6708

Error: 260005003 ICE CPU did not respond to polling within 180 seconds. Recon subsystem is presumed to be down. Some images may have been suspended. Thu Apr 29 06:05:11 2004 Error: 260005003 Host: fitzipct Process: procICE File: iceCpu.cxx Line: 343 ICE CPU did not respond to polling within 180 seconds. Recon subsystem is presumed to be down. Some images may have been suspended. Last ping was at Sun May 30 22:21:46 2032 with seq num 1818850152 iceConsole.log: Initializing System Memory (DRAM)... System Memory: 128MB, ECC Enabled (ECC-Memory Detected) PPC1-Bug

Error: 230006001 Process presumed down: vxmonitor-post Thu Jun 22 14:57:42 2006 Error: 230006001 Host: ls01 Process: procICE File: iceCpu.cxx Line: 360 Process presumed down: vxmonitor-post

Applies to

Models
LightSpeed
Product
LightSpeed 3.X (Ultra, Plus, QX/i) - Octane Always 4 or 8 slice MDAS 308i.2_H3.1M5 LightSpeed 3.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 308I.2_H3.1M5 305.4_H3.1M4 - obsolete 303.1_H3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2225004-9 73GB scan disk (replaces 2225004-8 and 18 GB p/n 2225004-4 and 36 GB scan disk 2225004-5) See Long Description for Details. Part #: 2225004-9 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: 73GB 15000 RPM Scan Data Drive FRU: Yes Status: Orderable Comments: Per ECR 2046173, 212008898, 8/8/07. ** FE ASSIST ** Refer to Service Note SN07110612a For Jumper information (12/3/07 212001261) Part #: 2225004-8 Part Type: GP CONSUMABLES Part Weight: 1.61 LBR Repairable: No Returnable: No Hazardous Material: N Description: HARD DRIVE 73GB 15KRPM SCAN DATA FRU: Yes Status: Orderable Comments: Added per ECR 2031409, 7/14/06, ms11892. ** FE ASSIST ** If no stock of the 2225004-8 order 2225004-9 as the replacement. Refer to Service Note SN07110612a (212001261 12/3/07)
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues