GE Healthcare · CTLightSpeedVCT26000151450004260001007260001012
LightSpeed 3.X (Ultra, Plus, QX/i) - GRE
Symptom
Scan aborts WITH Communications with the CCB sub-system has not been established. Scan aborts WITH Communications with the CCB sub-system has not been established
Error message
Error: 260001514 RCIB Bus: Fault Line Interrupt Occurred. Timeout Waiting for Host Response to Fault Line Verification. See long description for full error Wed Apr 27 01:00:51 2005 Error: 260001514 Host: orp Process: orp_subsys File: notiftsk.c Line: 1368 Running: System Reset Wed Apr 27 01:00:51 2005 Host : Das Ermes # : 260001514 Exception Class : Abort Severity : Secondary File : notiftsk.c Line# : 1368 Function : No System Function Reported Scan Type : None/Unknown Scan: 50004/1/0 RCIB Bus: Fault Line Interrupt Occurred. Timeout Waiting for Host Response to Fault Line Verification. (Timeout:500 msecs). Error: 260001007 CAN serial link error. A CC error was detected when sending a message to the CCB See the long description for the full error code Tue Nov 13 21:06:38 2007 Error: 260001007 Host: OBC Process: HOP File: hop_cc_send.c Line: 196 Function: No System Function Reported Scan: 2552/1/0 Type: None/Unknown Exception Level: Pri/Soft Time: 21:06:38:745 Log Series: 937 CAN serial link error. An CC error was detected when sending a message to the CCB. Message code: 1004 Node ID: 52 Object ID: 0 Send status: 5 Error: 260001012 The Rotating Controller Interface Bus (RCIB) fault
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 3.X (Ultra, Plus, QX/i) - GRE Always 4 or 8 slice MDAS. H2_P_M4 H2_P_M8 LightSpeed 3.X - GRE Method of visual identification Small rounded gantry similar in appearance to LS 2.X and 4.X. Always has an MDAS4 or MDAS8 Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da Always has an MDAS4 or MDAS8 List of Supported Console Types GOC3 Will have a DARC, SDDA and IG GOC4 Will have a DARC 2 and IG List of softwares the product can run 07MW11.10 SP3 06MW29.7 - obsolete 06MW03.5 - obsolete 05MW14.5 - obsolete 04MW44.9 - obsolete 04MW13.7 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2159578-3 collimator control board used on 1X/2X/3X/4X/5X See Long description for details Part #: 2159578-3 Part Type: GP CONSUMABLES Part Weight: 2.18 LBR Repairable: No Returnable: No Hazardous Material: N Description: COLLIMATOR CONTROL BOARD FRU: Yes Status: Orderable Comments: Not Available
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.