Pular para o conteúdo principal
GE Healthcare · CTLightSpeed26011828830585

LightSpeed 5.X (80/100kW)

Symptom

Scanner hardware stopped scan with KV errors. Scanner hardware stopped scan with Error: 260118215 (60-0318)kV did not reach 75 percent after 20ms.

Error message

Error: 260118288 (60-0332)too many kV_regul. number of kV_regul > number of KVRegulMax. (> 10 consecutives See LD for complete error Mon Sep 25 16:34:07 2006 Error: 260118288 Host: orp Process: orp_subsys File: gen_exec.c Line: 2376 Mon Sep 25 16:34:07 2006 Host : Jedi Ermes # : 260118288 Exception Class : Abort Severity : Primary File : gen_exec.c Line# : 2376 Function : No System Function Reported Scan Type : Scout Scan: 30585/1/1 (60-0332)too many kV_regul. number of kV_regul > number of KVRegulMax. (> 10 consecutives (On Db configurable)) Z6,Z7= 14h,0h

Applies to

Models
LightSpeed
Product
LightSpeed 5.X (80/100kW) LightSpeed 5.X (80/100kW) ~ Use with both MDAS and GDAS ~ GRE Console HP_H_G16 HP_H_M16 HP_O_M16 LightSpeed 5.X (100kW) GDAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2266521-5 HV Tank (replaces 2266521-4 and 2266521-3 - backward compatable) used on LS 7.X, LS 5.X and LightSpeed RT. See Long Description for details 2266521-5 HV Tank (replaces 2266521-4 and 2266521-3 - backward compatable) used on LS 7.X, LS 5.X and LightSpeed RT. Dash Tanks includes the EMI Shielding needed for G-Das Part #: 2266521-5 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: JEDI H4 MPHVTANK WITH SHIELDS FRU: Yes Status: Orderable Part #: 2266521-4 Part Type: GP CONSUMABLES Part Weight: -5499945.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: JEDI H4 MPHVTANK WITH SHIELDS FRU: Yes Status: Orderable Part #: 2266521-3 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 200.84 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: JEDI H4 MPHVTANK WITH SHIELDS FRU: Yes Status: Orderable Comments: added per ecr 2005924, as replacement for 2266521 and 266521-2. ms13452 040923 **If there is no Local Stock for the -3 part, you may order the -4 version which is backward compatible** ms0262 060815 *** 2ND OPINION NOTES *** Run the No-load KV test to verify HV tank is the error. D3194T PERFORMIX PRO-100 TUBE UNIT - Monopolar Hercules Tube used on 5.X (80/100), LS RT16, 7.X See Long Description for Details D3194T PERFORMIX PRO-100/VCT, RT16 and RT16 Xtra TUBE UNIT - Monopolar Hercules Tube used on 5.X (80/100), LS 5.X RT16, 7.X Part #: D3194T Part Type: GP TUBES Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: PERFORMIX PRO100/VCT (RG) & RT16 Systems FRU: Yes Status: Orderable Comments: Added per VCT RSPL, ms13452 041216 Comments: Added per VCT RSPL, ms13452 041216 FE ASSIST MSxxxx August 2005 D3194T = VCT / PRO 100 TUBE. VCT 32 or VCT 64 7.x systems may ONLY use D3194T tubes. 5.x HP 100 and HP80 Gdas systems may use either D3192T or D3194T tubes. '*** 2ND OPINION NOTES *** Perform Rotor Control Diagnostics--If tube rotor fails to reach RUN mode in three ACCEL steps look for: High Drive output currents Low Drive output voltages and Low Drive output frequencies. This indicates a load problem with tube. With rotor running turn off Gantry 120V power at PDU. Listen for excessive noise and short deceleration time. A good rotor will coast for several minutes without power. NOTE TO FES AND DCOS TEAM: IF SELLING TO AN END USER AND NOT INSTALLED BY A GE FE, AN END USER CERTIFICATION LETTER MUST BE SIGNED BY THE CUSTOMER. UPON RECEIPT OF SIGNED LETTER, ORDER CAN BE PLACED. **TO OBTAIN COPY OF CERT. LTR. CONTACT DCOS TEAM AT 800-558-2040 OPTION 2. **ANY QUESTIONS, CONTACT LYNN WESTBROOK OR ROD MCCRIMMON. ******* NOTE: D3194T tubes now manufactured, have a one inch longer high voltage cable. The longer cable is necessary for all Wide Bore Gantries. ( If a very old D3194T tube is installed on a Wide Bore Gantry, the high voltage cable may be too short, and a new D3194T tube must be ordered ).
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.