GE Healthcare · CTLightSpeedVCT200610002
LightSpeed 7.x (VCT, VCT Select, Pro32)
Symptom
ECG Trace fails to display on the console. See Long Description for more details. ECG Trace intermittently fails to display on the console, requiring a software reboot to recover. Problem seen after upgrading to 06MW03.4 software. The ECG trace works the first time after a reboot but after any interruption, the trace cannot be restored using the Gating BPM button.
Error message
200610002 Communication error between EkgBroker and EkgViewer See Long Discription for Details Wed Dec 9 08:43:19 2009 Error: 200610002 Host: vct1 Process: EKGBroker File: EKGBroker.cxx Line: 292 EKGBroker could not communicate with EKGViewer. Message: Communication error between EkgBroker and EkgViewer. EkgViewer needs to be restarted. Error: 5
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
- Component
- 5115334 Switch, Ethernet 8-port GBe (or equivalent 5270193) - used on GOC OC and HD gantry(5270193) See Long Description for Details. PN 5270193 will eventually replace 5115334. As per the ECR listed below the old switch power supply is no longer in production. When current stock is delpleted this part number will be the replacement. Part #: 5270193 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: Ethernet Switch with power supply and power cord. FRU: Yes Status: Orderable Comments: Per ECRs 2053301 & 2055972, 212008898, 4/7/08. Part #: 5115334 Part Type: GP CONSUMABLES Part Weight: 0.22 LBR Repairable: No Returnable: No Hazardous Material: N Description: Switch, Ethernet 8-port GBe FRU: Yes Status: Orderable
- Area
- CT
Full cause & step-by-step resolution