Pular para o conteúdo principal
GE Healthcare · CTHiSpeed18310250199

Scanning aborted with Tube temp / pressure switch open.

Symptom

Scanning aborted with Tube temp / pressure switch open - document in RFS - critical issue Scanning aborted with Tube temp / pressure switch open.

Error message

Error: 183102 Tube temperature/pressure switch open See the LD for the full error code Error: 183102 Tube temperature/pressure switch open Error: 183102 Host: OBC Process: Generator File: gen control.c 1.39 Line: 678 Function: XRAY GENERATION : Tube Pressure Sense Scan: 50199/1/4 Type: Axial Exception Level: Pri/Soft Time: 06:32:17:830 Log Series: 106 Tube temperature/pressure switch open.

Applies to

Models
HiSpeed
Product
HiSpeed CT/i 4.X HiSpeed CT/i 4.x with Zeus HSA Gantry and SGI Indigo Console HiSpeed CT/i with:HSA GANTRY CT/I (RHAPSODE) CONSOLE (Indigo Computer) Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
D3101T / D3102T X-ray Tube Used In CT/i and HiSpeed Advantage Part #: D3102T Part Type: GP TUBES Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: TUBE HSA/CTI FRU: Yes Status: Orderable Comments: FLS - NEW 1994 TUBE CAT. FOR RELOAD REPLACES D1199ZR (RELOADED) .9 X .7FS, 1.2-1.2 FS HISPEED ADVANTAGE DIMS 37 X 27 X 28 (LWH) MAXIRAY EQUIVALENT IS D3109T TUBE FAN= 46-219465P1 AM SOURCE = GEMS-MM (MEXICO) *** FLS Tube Names Systems Compatible Heat Capacity Focal Spot Catalog # Model # Insert # Casing # Tube Unit # HSA(Zeus) HiSpeed CT/I, or HiSpeed Advantage 3.5MHu 0.9x0.7/1.2x1.2 D3102T MX165 46-309300G2 46-309500G2 46-309500G801 Tubes started shipping half machined, half cast casings in early 2005. This was stated to be a tempory change in process while a new vendor was being evaluated.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues