Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT260112009244634

LightSpeed 7.x (VCT, VCT Select, Pro32)

Symptom

Attention: The Scanning Hardware is not available. Please see long description. Attention: The Scanning Hardware is not available, please wait for the hardware reset successfull status posted in the feature status area on the right monitor. If the hardware reset fails, then reset the scan hardware from the Service Desktop System Resets button.

Error message

Error: 260112009 Emergency Off button has been pressed See LD for full error Fri May 20 15:06:56 2005 Error: 260112009 Host: tgp Process: tgp_subsys File: uif_table_tilt_mgr.c Line: 640 Fri May 20 15:06:56 2005 Host : Table Gantry Processor Ermes # : 260112009 Exception Class : Operator Abort Severity : Secondary File : uif_table_tilt_mgr.c Line# : 640 Function : SYSTEM POWER CONTROL : Emergency Stop Scan Type : None/Unknown Scan: 0/0/0 Emergency Off button has been pressed

Error: 244634 Subsystem State; DCB:up, CCB:up, -TCB:down -, JEDI:up, TGP/ORP:up TCB:DOWN See Long Description Sun Aug 29 11:13:58 2004 Error: 244634 Host: erct Process: fwmgr File: fwmgr utils.c Line: 660 Function: No system function reported Scan Type: Not scanning Scan: 0/0/0 Scan Seq Id: 0 Subsystem State; DCB:up CCB:up TCB:down JEDI:up TGP/ORP:up.

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues