Pular para o conteúdo principal
GE Healthcare · CTHiSpeed26000650050000

Scanner Hardware Stopped Scan.

Symptom

Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan

Error message

Error: 260006500 The axial AC Drive reported fault code: F4 - Undervoltage Fault See the LD for the full error code Error: 260006500 The axial AC Drive reported fault code: F4 - Undervoltage Fault Tue Nov 16 13:47:02 2004 Error: 260006500 Host: STC Process: Axial File: AX2fault.c Line: 196 Function: AXIAL CONTROL : axial control Scan: 50000/2/0 Type: None/Unknown Exception Level: Pri/Most Severe Time: 13:47:03:015 Log Series: 117 The axial AC Drive reported fault code: F4 - Undervoltage Fault The Drive internal DC bus voltage dropped below 325V. Possible low voltage condition on 480 VAC in gantry or power interruption. Also, may indicate excessive run or braking power required due to sluggish Motor.

Applies to

Models
HiSpeed
Product
HiSpeed QX/i - GOC2 103L.5_HSQXIM4 103L.4_HSQXIM4 102L.1_HSQXIM3 HiSpeed QX/i - GOC2 Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types GOC2 List of softwares the product can run 103L.5_HSQXIM4 103L.4_HSQXIM4 - obsolete 102L.1_HSQXIM3 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
46-170021P15 (250V,8A) Fuse used in axial dynamic brake circuit as CT2A1A10F1 and found in FAN ASSY- GANTRY, and F4, F5 on HEMRC Interface Board and other places See long description for details Part #: 46-170021P15 Part Type: GP CONSUMABLES Part Weight: 0.45 LBR Repairable: No Returnable: No Hazardous Material: N Description: 008.000A 0250V 3AB SLOW UL/ FRU: Yes Status: Orderable Comments: FLS - FUSE 8 AMP - 250 VOLT 3AB
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues