GE Healthcare · CTLightSpeed260001007260001012
LightSpeed 4.X - Octane
Symptom
Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan
Error message
Error: 260001007 CAN serial link error. A CC error was detected when sending a message to the CCB See the long description for the full error code Tue Nov 13 21:06:38 2007 Error: 260001007 Host: OBC Process: HOP File: hop_cc_send.c Line: 196 Function: No System Function Reported Scan: 2552/1/0 Type: None/Unknown Exception Level: Pri/Soft Time: 21:06:38:745 Log Series: 937 CAN serial link error. An CC error was detected when sending a message to the CCB. Message code: 1004 Node ID: 52 Object ID: 0 Send status: 5 Error: 260001012 The Rotating Controller Interface Bus (RCIB) fault line has been asserted See Long Description for the full error code Sat Jun 9 14:37:06 2001 Error: 260001012 Host: OBC Process: HOP File: hop_fault_state.c Line: 137 Function: No System Function Reported Scan: 9553/3/1 Type: Axial Exception Level: Pri/Most Severe Time: 14:37:06:030 Log Series: 366 The Rotating Controller Interface Bus (RCIB) fault line has been asserted. None of the sub-systems have reported a problem. Check nodes or fault line for failure.
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 4.X - Octane sw_version: 405I.2_H4.0M5 LightSpeed 4.X - Octane Gantry Model # 2339985 Console Model #'s 2341923 and 2341923-2(D3D) check_config shows LightSpeed 16 / LightSpeed 16 Product: LightSpeed16 sw_version: 405I.2_H4.0M5 or newer machine_type: Octane host_id: 3232297858 tube_type: 15 Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2159578-3 collimator control board used on 1X/2X/3X/4X/5X See Long description for details Part #: 2159578-3 Part Type: GP CONSUMABLES Part Weight: 2.18 LBR Repairable: No Returnable: No Hazardous Material: N Description: COLLIMATOR CONTROL BOARD FRU: Yes Status: Orderable Comments: Not Available
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.