GE Healthcare · CTLightSpeed26000680650004260006804
LightSpeed 3.X (Ultra, Plus, QX/i) - GOC2
Symptom
Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan
Error message
Error: 260006806 Axial Motor Control (AMC) logic will NOT accept commands without a Valid Configuration/Characterization file. See the Long Description for the full error code Thu Nov 18 20:34:37 2004 Error: 260006806 Host: STC Process: Axial File: amc_exec.c Line: 3464 Function: No System Function Reported Scan: 50004/1/0 Type: None/Unknown Exception Level: Pri/Most Severe Time: 20:34:37:390 Log Series: 37 Axial Motor Control (AMC) logic will NOT accept commands without a Valid Configuration/Characterization file. (Cmd: 0x1804ffff). Error: 260006804 Axial Motor Control (AMC) logic could not configure the AXIAL drive via the CAN network during configuration See long description for complete error Thu Nov 18 20:34:37 2004 Error: 260006804 Host: STC Process: Axial File: amc_exec.c Line: 3606 Function: AXIAL CONTROL : axial communications Scan: 50004/1/0 Type: None/Unknown Exception Level: Secondary Time: 20:34:37:420 Log Series: 37 Axial Motor Control (AMC) logic could not configure the AXIAL drive via the CAN network during configuration. Reset scan hardware to retry.
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 3.X (Ultra, Plus, QX/i) - GOC2 This is HP CPU host running LINUX with Pegasus IG. 308L.8_H3.3M4 ~ 4 or 8 slice MDAS. ~ Could be an Ultra, Plus or QXi system. Must run swokinstall -p to verify. ~ See Long description for details . LightSpeed 3.X MDAS4 Gantry and GOC2 Gantry - 2281177-2, DAS - 2272768, Console – 2341104-2, Table - 2276776, PDU - 2269902 Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 46-170021P15 (250V,8A) Fuse used in axial dynamic brake circuit as CT2A1A10F1 and found in FAN ASSY- GANTRY, and F4, F5 on HEMRC Interface Board and other places See long description for details Part #: 46-170021P15 Part Type: GP CONSUMABLES Part Weight: 0.45 LBR Repairable: No Returnable: No Hazardous Material: N Description: 008.000A 0250V 3AB SLOW UL/ FRU: Yes Status: Orderable Comments: FLS - FUSE 8 AMP - 250 VOLT 3AB
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.