GE Healthcare · CTLightSpeedVCT260007768245207
LightSpeed 5.X (80/100kW)
Symptom
Can't scan with Pink Warning: Software has detected an invalid DAS configuration.
Error message
Error: 260007768 DAS Converter Boards are not installed in valid configuration See LD for complete error Mon Jan 23 07:02:45 2006 Error: 260007768 Host: DAS Process: DCB Firmware File: cnvmgr.c Line: 197 Function: DATA ACQUISITION : Scan Data Acquisition Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Soft Time: 07:02:445:750 Log Series: 30 DAS Converter Boards are not installed in valid configuration. The Following Card List Indicates Potential Missing Boards Error: 245207 Software has detected an invalid or mistached DAS type. This will prevent the system from scanning. see long description Mon Apr 3 08:29:12 2006 Error: 245207 ls16 scanmgr validate request.cxx 378 Software has detected an invalid or mistached DAS type. This will prevent the system from scanning.
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 5.X (80/100kW) LightSpeed 5.X (80/100kW) ~ Use with both MDAS and GDAS ~ GRE Console HP_H_G16 HP_H_M16 HP_O_M16 LightSpeed 5.X (100kW) GDAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Area
- CT