Pular para o conteúdo principal
GE Healthcare · CTLightSpeed183327219800

LightSpeed 4.X - GRE

Symptom

Scanner Hardware Stopped Scan HEMRC Errors See Long Description for more information. Scanner Hardware Stopped Scan message appears during Helical Scanning with the following Errors: 185553 (The HEMRC drive has detected a Line Loss - Alarm status: 20H. This occurs when the input power to the drive falls below 85% of the nominal Bus voltage) Possible causes: X-Ray tube stator , HEM-IT, HEMRC power supply 183318 (The HEMRC output current is out of 45.0 percent WARNING tolerance), Monitored state = 2 (2=Run state) Expected current = 2.748 A Actual current = 4.097 A. 183319 (The HEMRC output current is out of 55.0 percent MAXIMUM tolerance), Monitored state = 2 (2=Run state) Expected current = 2.748 A Actual current = 4.653 A. 183316 (The HEMRC failed to reach the Run speed in the allotted time), and 185548 (The HEMRC operating frequency is below the minimum value) Actual frequency: 106.973 Minimum frequency limit: 130.000 At Frequency signal: Inactive Expected: Inactive Rotor State = Transition

Error message

Error: 183327 HEMRC fault detected. Long Description contains complete error. Mon Jul 24 08:41:32 2006 Error: 183327 Host: OBC Process: Generator File: gen_abort.c 1.31 Line: 625 Function: XRAY GENERATION : Tube Rotor Control Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Most Severe Time: 08:41:28:555 Log Series: 31 HEMRC fault detected. See system log for additional error information. Fault latch address = ffb82b hex, Bit = D0

Error: 219800 The HEMRC AC Drive reported fault code: F12 - Overcurrent Fault See long description for details Error: 219800 The HEMRC AC Drive reported fault code: F12 - Overcurrent Fault Wed Dec 23 18:05:18 2009 Error: 219800 Host: OBC Process: Generator File: HEMRCfault.c Line: 194 Function: XRAY GENERATION : Tube Rotor Control Scan: 1002/1/1 Type: Scout Exception Level: Pri/Most Severe Time: 18:05:17:765 Log Series: 79 The HEMRC AC Drive reported fault code: F12 - Overcurrent Fault Check for a short circuit at the drive output or excessive load conditions at the motor.

Applies to

Models
LightSpeed
Product
LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2115199 HEMRC - HIGH EFF MTR ROTOR CTRL used on LightSpeed 2.x (Plus), 3.x (Ultra) and 4.x (16 Slice) Systems See Long Description for more info and picture Part #: 2115199 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 7.28 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: HIGH EFF MTR ROTOR CTRL FRU: Yes Status: Orderable Comments: FLS - HIGH EFF MTR ROTOR CTRL PURCH SPEC USED ON FAST RHAPSODE 7/1/97 M3 FW28 GWENN 08992 THIS IS THE AC DRIVE FOR HEMRC THIS IS OUT-SOURCED TO A 3RD PARTY FOR ALLEN-BRADLEY. NOT DO NOT DISCOUNT - PER SHAKUN 10/30/97 GWENN MS08992 6/02 BOT FROM HOLT ELECTRIC A/B P#: 1336SBRF50-AN-EN901-MX1-L4 FLS - PLEASE CONVERT SPH ORDER TO OVERNIGHT SHIPMENT IF PART AVAILABLE IN MEMPHIS AND IF FEDEX SERVICE (INCLUDING 8:00 AM SERVICE) MEETS FE NEED TIME CRITERIA
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.