GE Healthcare · CTLightSpeed18221120525750009085218221360046
LightSpeed 4.X - GRE
Symptom
Cradle will not drive in or out. Cradle will not drive in or out
Error message
Error: 182211 Velocity error detected by hardware (LM628 excessive position error). See the Long Description for the full error Tue Aug 29 08:09:30 2006 Error: 182211 Host: ETC Process: Cradle File: cmcLm628IsrTsk.c 1.19 10/16/9 Line: 185 Function: PATIENT ANATOMY POSITIONING : Cradle Drive Scan: 20525/2/1 Type: Helical Exception Level: Pri/Soft Time: 08:09:30:260 Log Series: 595 HW error: Velocity error detected by hardware (LM628 excessive position error). Expected Velocity 75000 (.001 mm/sec) Actual Velocity 0 (.01 mm/sec) Pos Err Tol 2540 (.01 mm) Cur Pos 90852 (.01 mm) Cradle Dir: 1 (0 = Out 1 = In [Gantry]). Error: 182213 Velocity tracking error See Long Description for the full error code Tue Mar 15 15:28:05 2005 Error: 182213 Host: ETC Process: Cradle File: CmdMotion.c 1.33 7/23/96 Line: 92 Function: PATIENT ANATOMY POSITIONING : Cradle Drive Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Soft Time: 15:28:05:770 Log Series: 37 HW error: Velocity tracking error. Expected Velocity 500 (.001 mm/sec) Actual Velocity 9382 (.01 mm/sec) Current Position 60046 (.01 mm) Velocity Tolerance 250 (.001 mm/sec) Cradle Dir: 1 (0 = Out 1 = In [Gantry]). Error: 184165 Cradle was not
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 46-264368G3 ETC board See long description for details Part #: 46-264368G3 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: ENHANCED TABLE CONTROL HELIOS 2 FRU: Yes Status: Orderable Comments: Not Available 46-264370G1 CRADLE AMP used on Zeus table (1.X-4.X) See long description for details Part #: 46-264370G1 Part Type: GP CONSUMABLES Part Weight: 1.81 LBR Repairable: No Returnable: No Hazardous Material: N Description: CRADLE AMP FRU: Yes Status: Orderable Comments: FLS - B: IF ORDERING 46-264370G1 FOR TROUBLESHOOTING A CRADLE THAT IS EXHIBITING ERRATIC JERKY MOTIONS, ORDER THE 2120791-9 RELAY (24 V POTTER &BRUMFIELD). MSA6521 5/29/01 CRADLE AMP BD 2/8/01- T3202CE DELISTED DUE TO LOW ICV. NO POOL ADDS. MS12464 FLS - PLEASE CONVERT SPH ORDER TO OVERNIGHT SHIPMENT IF PART AVAILABLE IN MEMPHIS AND IF FEDEX SERVICE (INCLUDING 8:00 AM SERVICE) MEETS FE NEED TIME CRITERIA 2269169 CRADLE DRIVE ASM FOR CATHAY TABLE. See Long Description for details Part #: 2269169 Part Type: GP REPAIRABLES Part Weight: 14.23 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: CRADLE DRIVE ASM FOR CATHAY TABLE FRU: Yes Status: Orderable Comments: FLS - CRADLE DRIVE ASM FOR CATHAY TABLE. P/N 2269169 REPLACES P/N 46-296235G2 PER PCN 220590USE FIRST EXISTING STOCK. MS20544 ON 8/23/02.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.