Symptom

Fastcal is failing with beam obstructed errors. Fastcal is failing with beam obstructed errors

Error message

Error: 230017082 Channel in non-bowtie airscan out of range - row 0, channel 10, value = 4.3 The row number can be VERY confusing. Please see Long Description for how to decode the ROW NUMBER Tue Mar 22 20:12:04 2005 Error: 230017082 Host: ct01 Process: accs File: aircal_processing.c Line: 917 Channel in non-bowtie airscan out of range - row 0, channel 10, value = 4.394680 Click Here to Decode the ROW NUMBER

Error: 230017078 Aircal failed to process scan See Long Description for complete error Tue Mar 22 20:12:04 2005 Error: 230017078 Host: ct01 Process: accs File: accs_events.cxx Line: 463 Aircal failed to process scan - Exam = 50342, Series = 1, Scan = 1, Seconds = 1111522307 Microseconds = 108580 View scanfileto check data and header information

Error: 230023046 Calibration limit check failed. Beam may be obstructed See LD for complete error Tue Mar 22 20:12:04 2005 Error: 230023046 Host: ct01 Process: dailyPrepRx File: DailyPrepUtils.c Line: 3953 Calibration limit check failed. Beam may be obstructed

Applies to

Models
LightSpeed
Product
LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
5331736 - Helios Filter Asselmbly FRU (replaces 2291958) Used on Helios 3 and 4 Collimators. See Long Description for more Details Part #: 5331736 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: Helios Filter Asselmbly FRU FRU: Yes Status: OrderableComments: Per ECR 2082045, 212008898, 6/16/09. Part #: 2126674 Part Type: GP CONSUMABLES Part Weight: -144548.55 LBR Repairable: No Returnable: No Hazardous Material: N Description: FILTER ASSEMBLY, HELIOS COLLIMATOR FRU: NO Status: Non Orderable
Area
CT

Full cause & step-by-step resolution