GE Healthcare · CTLightSpeed182207104947063665430
LightSpeed 4.X - GRE
Symptom
Any time the table moved flashing eights would show on gantry display and table loses reference Any time the table moved up or down flashing eights would show on gantry display and table loses reference
Error message
Error: 182207 HW error: Alignment error between Encoder and Potentiometer See Long Description for complete error Tue Aug 5 15:05:39 1997 Error: 182207 Host: ETC Proc:Cradle File: cmcCorrErr.c 1.22 3/11/92 Line: 51 Function: PATIENT ANATOMY POSITIONING : Cradle Drive Scan: 10494/1/0 Type: None/Unknown Exception Level: Pri/Soft Time: 15:05:39:320 Log Series: 28 HW error: Alignment error between Encoder and Potentiometer. Potentioment Position 70636 (.01 mm) Encoder Position 65430 (.01 mm) Tolerance 2540 (.01 mm)
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2271256 SIGNAL HARNESS FOR H2 TABLE See Long Description for part details 2271256 SIGNAL HARNESS FOR H2 TAB Part #: 2271256 Part Type: GP CONSUMABLES Part Weight: 1.09 LBR Repairable: No Returnable: No Hazardous Material: N Description: SIGNAL HARNESS FOR H2 TABLE FRU: Yes Status: Orderable Comments: FLS - THIS IS SIGNAL HARNESS USED ON LIGHTSPEED H2/FRU CODE GPCP/ FOUND NO STOCK/PDR ALREADY IN PLACE/MS6255 01/19/02 ASIA SOURCED PART
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.