Pular para o conteúdo principal
GE Healthcare · CTHiSpeed

HiSpeed CT/i 4.X with Indigo and HSA Gantry

Symptom

CTi: 4.4, 4.5 IRIX fails to boot intermittently after software load and may generate a loud screeching sound from the intercom See Long Description for additional detail CTi: 4.4, 4.5 IRIX fails to boot intermittently after software load and may generate a loud screeching sound from the intercom. CTi 4.4 Indigo CPU. OC system Disk failed and was replaced. After the software load, the system would intermittently fail to come up to applications level. The message in the OC SYSLOG was: savecore GIO error / addr 0x400: 0x4083800 PANIC: IRIX KILLED DUE TO BUS ERROR @ PC:0xc0674ff8 ep:0xffffca60 The system may boot to applications if tried as many as 3-5 times. Once up in applications, the system would scan without error. If the system was shut down, it might take 3-5 times of rebooting before it would get to applications without error.

Error message

/usr/adm/SYSLOG: savecore: reboot after panic: PANIC: IRIX Killed due to Bus Error /usr/adm/SYSLOG: savecore: reboot after panic: <0>PANIC: IRIX Killed due to Bus Error

Applies to

Models
HiSpeed
Product
HiSpeed CT/i 4.X HiSpeed CT/i 4.x with Zeus HSA Gantry and SGI Indigo Console HiSpeed CT/i with:HSA GANTRY CT/I (RHAPSODE) CONSOLE (Indigo Computer) Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Software
4.4.19.1 for HiSpeed CT/i with with CT/I (RHAPSODE) CONSOLE (Indigo) - showprods returns Rhapsode Product 4.4.19.1 4.4.19.1 for HiSpeed CT/i with with CT/I (RHAPSODE) CONSOLE (Indigo) - showprods returns Rhapsode Product 4.4.19.1 Intro. Date 10/18/00 Mandatory FMI 25268 Rhapsode Product 4.4.19.1 software for HiSpeed CT/i with SGI Indigo computer Intro. Date 10/18/00 Mandatory FMI 25268 Application Software Set Set Part No. . 2283723-3 Apps. CD Part No 2282440-3 Apps. Ver. 4.4.19M4 Apps. CD Label HiSpeed CT/I App SW showprods/Service Desktop Rhapsode Product 4.4.19.1 Class C CD Part No. Ver. CD Label 2282441-2 4.419M4 HiSpeed CT/I Class C Software Class M CD Part No. Ver. CD Label 2282442-2 4.419M4 HiSpeed CT/I Class M Software Patch Info: P/N 2318818, 4.5 Patch Rhapsode Product 4.4.19.1 Rhapsode patch 4.5.3 ver 1 11/21/2001 Optional FMI 25287A Also available via AutoSC ,_,Number:,2283723-3_______, ,BOM,Rev:,___,PI Rev:,___,CR Nb/Typ:,_______,___, ,CTZ,US, ,JAROSCH,ANN,K , ,Rev:,000,PI Rev:,000,CR Nb/Typ:, , , ,Description/Name:,E,HISPEED CT/I OPER SFWR VER 4.4.19 M4, ,Date:,2000-11-13, ,AIL:, , ,R,S,S,F, ,0001,/,0002, ,Mark, ,Number, ,Quantity,,U/M,U,T,P,C,Description/Name, ,__,1 ,2282439-2 , 1.000,PC ,S,N,N,N,IRIX 6.2M4 OS FOR 4.4 SOF, ,__,2 ,2282440-3 , 1.000,PC ,S,N,1,1,HISPEED CT/I APP SFWR VER, , , , , , , , , , , ,
Component
2295259 : SOFTWARE PATCH FOR MEMORY UPGRADE See Long Description for Comments 2295259 : SOFTWARE PATCH FOR MEMORY UPGRADE Part #: 2295259 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: SOFTWARE PATCH FOR MEMORY UPGRADE FRU: Yes Status: Orderable Comments: FLS - CT/I 4.4 IRIX 384 MEMORY UPGRADE PATCH. VENDOR IS HM PRODUCT SOLUTIONS. MS26561 DSM 4/21/01
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.