GE Healthcare · CTHiSpeed18333819507
HiSpeed CTi 6.X with Octane and HSA Gantry
Symptom
Rotor does not spin up and HVDC Voltage has dropped below the operating level. See long description Rotor does not spin up and the HVDC Voltage has dropped below the operating level.
Error message
Error: 183338 The maximum allowable HVDC Rail voltage drop has been exceeded. See LD for complete error Tue Jul 25 09:43:35 2006 Error: 183338 Host: OBC Process: Generator File: gen_power.c 1.13 Line: 258 Function: SYSTEM MONITORING : HVDC Monitor Scan: 19507/1/1 Type: Scout Exception Level: Pri/Most Severe Time: 09:43:35:145 Log Series: 528 The maximum allowable HVDC Rail voltage drop has been exceeded. No load HVDC value: 663V 1st HVDC sample: 27V HVDC 15mS later: 27V Rotor LOV bit value: 1 From Feedback If error indicates approximately half of proper HVDC supply, this indicates strong chance of blown 700V fuse F1 or F2 on HEMRC I/F board in HEMRC
Applies to
- Models
- HiSpeed
- Product
- HiSpeed CTi 6.X HiSpeed CTi 6.X with Octane and HSA Gantry HiSpeed CTi 6.X with Octane and HSA Gantry Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2133533 CT Compact Power Distribution Unit 2133533 CT Compact Power Distribution Unit
- Area
- CT