Symptom

Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan

Error message

Error: 185557- hemrc serial link is down. Hemrc_+12V failure detected. Possible causes: HEMRC fuse (F5), Interconnects, HEMRC AC Drive, HEMRC Control Board See Long Description for complete error Mon Jul 24 08:41:32 2006 Error: 185557 Host: OBC Process: Generator File: scanbusDriver.c 1.5 Line: 243 Function: XRAY GENERATION : Tube Rotor Control Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Most Severe Time: 08:41:28:915 Log Series: 31 Could not send HEMRC message in the allotted time. HEMRC serial link is down. HEMRC_ISO_+12V Failure detected. Possible causes: HEMRC fuse (F5), Interconnects, HEMRC AC Drive,

Applies to

Models
LightSpeed
Product
LightSpeed 4.X - Octane sw_version: 405I.2_H4.0M5 LightSpeed 4.X - Octane Gantry Model # 2339985 Console Model #'s 2341923 and 2341923-2(D3D) check_config shows LightSpeed 16 / LightSpeed 16 Product: LightSpeed16 sw_version: 405I.2_H4.0M5 or newer machine_type: Octane host_id: 3232297858 tube_type: 15 Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
46-170021P15 (250V,8A) Fuse used in axial dynamic brake circuit as CT2A1A10F1 and found in FAN ASSY- GANTRY, and F4, F5 on HEMRC Interface Board and other places See long description for details Part #: 46-170021P15 Part Type: GP CONSUMABLES Part Weight: 0.45 LBR Repairable: No Returnable: No Hazardous Material: N Description: 008.000A 0250V 3AB SLOW UL/ FRU: Yes Status: Orderable Comments: FLS - FUSE 8 AMP - 250 VOLT 3AB
Area
CT