Pular para o conteúdo principal
GE Healthcare · CTHiSpeed2000004372000004232687225588

HiSpeed QX/i - Octane

Symptom

Images fail to display and get suspended in the recon queue

Error message

Error: 9056 connection to Netserver failed, exiting... See the long description for the full error code Thu Jun 23 07:02:15 2005 Error: 9056 Host: gtct3 Process: dbrserver File: server.c Line: 52 connection to Netserver failed, exiting...

Error: 200000437 Image installation failed for image (for some Exam/Series/Image number) Open Long Description for full error varaiations Sat Jul 16 13:11:02 2005 Error: 200000437 Host: hsa2 Process: dms File: InstallImage.cxx Line: 316 Image installation failed for image 7297/1/1. Reason: Response from SdC timeout. Thu Feb 2 13:32:10 2006 Error: 200000437 Host: ct01 Process: dms File: InstallImage.cxx Line: 713 Image installation failed for image 2336/999/1. Reason: DMS internal error occurred. image path = %s Error code: 200000423

Error: 9071 get mail failed in process_mail() See the long description for the full error code Thu Jun 23 07:02:15 2005 Error: 9071 Host: gtct3 Process: dbrserver File: proc.c Line: 376 get mail failed in process_mail()

Error: 1031 stored procedure failed See Long Description for Details. Sun Nov 14 20:00:37 2004 Error: 1031 Host: mcas Process: dbrserver stored procedure failed : -284 - getexamuidinfo failed Also ch

Applies to

Models
HiSpeed
Product
HiSpeed QX/i - Octane Always MDAS4 101I.2_HSQXIM5 101.4_HSQXIM4 HiSpeed QX/i - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 101I.2_HSQXIM5 101.4_HSQXIM4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Software
101I.2_HSQXIM5 See LD for ClassM & C 101I.2_HSQXIM5 used on HiSpeed (YMS) only Class M software - 2395173 or 2401759 (6 pack) Class C software - 2395172 or 2401759 (6 pack) Patch G.6 Mandatory FMI# 25363 (P/N 5136941)
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues