Symptom

Will not scan with "Not enough image space exists. Please remove images". See Long Description for screen capture. Scanner would not allow scouts, axial or helical scanning with message on left screen: Not enough image space exists. Please remove images.

Error message

Error: 200502005 eeeee/s/i not found in image management, cannot reallocate image space See Long Description for sample error Tue Oct 19 09:22:20 2004 Error: 200502005 Host: dgct Process: scanRx File: srx_sma_manage_image_space.c Line: 951 Function: No system function reported DGCT 10989/4/0 not found in image management, cannot reallocate image space

Applies to

Models
LightSpeed
Product
LightSpeed RT16 & RT16 Xtra and Vision WB_H_G16_G_GTL = RT16, Xtra WB_H_G16_G_GTL650 = Vision WB_H_G16_G_GTM = RT16, Xtra or Vision LightSpeed RT16 & RT16 Xtra with GRE and GDAS16 gantry Links to the Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Software
08BW17.7 For LightSpeed RT16 & RT16 Xtra See LD for details 08BW17.7 For LightSpeed RT16 & RT16 Xtra sw_hw_string: 08BW17.7.WB_H_G16_G_GTM Released with FMI 22918 Purpose This FMI delivers software release 08BW17.7 to LightSpeed RT16 and LightSpeed Xtra. This software improves the tube warm-up procedure, and delivers safety and quality fixes to improve system reliability. See Product Quality Fixes and Product Safety Fixes section for further details. See Appendix C (In attachment) for improved tube warm-up steps. Product Quality and Safety fixes are the fixes that were included in Service Pack 2.0 for 07MW11.0, 07MW18.4, and 07BW08.4. 5315090 - Class C (Tab# 182) 5315091 - Class M (Tab# 182)
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.