Symptom

"Scan hardware Stopped Scan" with collimator errors. "Scan hardware Stopped Scan" with collimator errors

Error message

Error: 260003164 A Z-Axis Update Contained a Move That Is Invalid For CAM A. Illegal distance for CAM A encoder count conversion. Illegal distance for CAM B encoder count conversion. See LD for complete error Fri Oct 12 12:43:36 2001 Error: 260003164 Host: HSC Process: CCB Firmware File: aperture_z_axis.c Line: 230 Function: XRAY GENERATION : Xray Collimation and Filtration Scan: 50267/1/0 Type: None/Unknown Exception Level: Pri/Soft Time: 12:43:36:745 Log Series: 57 A Z-Axis Update Contained a Move That Is Invalid For CAM A. Position : 15799 Sequence Number: 0 Mon Oct 2 10:06:25 2006 Error: 260003164 Host: HSC Process: CCB Firmware File: aperture_z_axis.c Line: 230 Function: XRAY GENERATION : Xray Collimation and Filtration Scan: 50646/4/0 Type: None/Unknown Exception Level: Pri/Soft Time: 10:06:25:445 Log Series: 898 A Z-Axis Update Contained a Move That Is Invalid For CAM B. Position : -259 Sequence Number: 38

Applies to

Models
LightSpeed
Product
LightSpeed RT16 & RT16 Xtra and Vision WB_H_G16_G_GTL = RT16, Xtra WB_H_G16_G_GTL650 = Vision WB_H_G16_G_GTM = RT16, Xtra or Vision LightSpeed RT16 & RT16 Xtra with GRE and GDAS16 gantry Links to the Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2159578-3 collimator control board used on 1X/2X/3X/4X/5X See Long description for details Part #: 2159578-3 Part Type: GP CONSUMABLES Part Weight: 2.18 LBR Repairable: No Returnable: No Hazardous Material: N Description: COLLIMATOR CONTROL BOARD FRU: Yes Status: Orderable Comments: Not Available
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.