Pular para o conteúdo principal
GE Healthcare · CTLightSpeed290009010290009016290009013290012004

LightSpeed 5.X (80/100kW)

Symptom

Image recon is failing intermittently - document in RFS - critical issue See long description for details. Image recon is failing intermittently.

Error message

Error: 290009010 Image suspended due to: image installed with status -1. See long description for full error Fri Sep 30 15:00:17 2005 Error: 290009010 Host: darc Process: recon_control File: ImageSetRequest.cxx Line: 4129 Function: Image Generation : SBC Processing Job Type: RECON_JOB_PROS_AXIAL_RECON Scan: 613/2/6 Image: 613/2/22 Image suspended due to: image installed with status -1. Number of retry so far: 2. IG node: 0.

Error: 290009016 Recon is going to SHUTDOWN mode due to XXXXXX. See Long Description for Error variation Tue Jul 12 14:31:49 2005 290009016 4 darc recon_control recon_control_cb.cxx 464 Recon is going to SHUTDOWN mode due to no IG can be used to perform reconstruction. Mon Aug 31 23:13:32 2009 Error: 290009016 Host: darc Process: recon_control File: ImageSetRequest.cxx Line: 6029 Recon is going to SHUTDOWN mode due to maximum number of consecutive suspends reached.

Error: 290009013 Not restarting IG from fd: 31 because we have reached the max number of consecutive resets. See LD for full Error Tue Jul 12 14:31:49 2005 290009013 4 darc recon_control recon_control_cb.cxx 460 Not restarting IG from fd: 31 because we have reached the max number of consecutive rese

Applies to

Models
LightSpeed
Product
LightSpeed 5.X (80/100kW) LightSpeed 5.X (80/100kW) ~ Use with both MDAS and GDAS ~ GRE Console HP_H_G16 HP_H_M16 HP_O_M16 LightSpeed 5.X (100kW) GDAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2362872 Westville IG Module used on GOC3 and GOC4 consoles. See Long Description for details. 2362872 Westville IG Module used on GOC3 and GOC4 consoles. Part #: 2362872 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 19.84 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: IMAGE GENERATION COMPUTER - GRE FRU: Yes Status: Orderable Comments: **FE Assist** applies to CT Not Pet systems. If no stock is available order a Jarrell IG 5147443 to replace a Westville IG (2362872), you MUST order p/n 5193989 software release 06MW29.7, and perform a LFC. See Service note SN06101010A. PET must receive the Westville IG.(12/12/06 212001261) *** 2ND OPINION NOTES *** See the following URL for additional command line diagnostic troubleshooting: http://gemsvst4p.med.ge.com:7005/facets/CT/GRE%20command%20line%20diag%20test.pdf?
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.