GE Healthcare · CT182405100642072920807184836

Lightspeed 1.X - Octane

Symptom

Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan

Error message

Error: 182405 HW/SW error: Uncommanded elevation motion. See long description for full error Wed Mar 19 13:22:41 2008 Error: 182405 Host: ETC Process: Elevation File: unComMotion.c 1.28 11/5/92 Line: 170 Function: PATIENT ANATOMY POSITIONING : Elevation - Tilt Scan: 10064/2/6 Type: Axial Exception Level: Pri/Soft Time: 13:22:42:215 Log Series: 54 HW/SW error: Uncommanded elevation motion. Current Position 20729 (.01 mm) Last Stationary Position 20807 (.01 mm) Position Tolerance 75 (.01 mm) Current Encoder cnts 3446 Last Stationary Encoder Cnts 3435

Error: 184836 motor controller detects uncommanded motion. Aborting any active scan Tue Aug 10 19:04:09 2004 Error: 184836 Host: ETC Process: Elevation File: elev scan.c 1.17 9/30/92 Line: 178 Function: No System Function Reported Scan: 4530/2/7 Type: Helical Exception Level: Pri/Most Severe Time: 19:04:08:905 Log Series: 376 motor controller detects uncommanded motion. Aborting any active scan

Applies to

Product
Lightspeed 1.X - Octane 10.5_2.8.2I_H1.3M4 10.4.1_2.8.2I_H1.3M4 - obsolete Lightspeed 1.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config showprods|grep Apps note: if check_config does not work then the software is not up to date with all FMI's. Check FMI history and verify that all patches have been loaded since last software reload. Ref FMI 25269 List of Supported Console Types Octane with Pegasus Recon hardware. may have an 02 SBC - Run df -k to see if the sbc is partitioned. List of softwares the product can run 10.5_2.8.2I_H1.3M4 10.4.1_2.8.2I_H1.3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3176T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2269175-3 LONGITUDINAL ENCODER ASM (Used on HSA, CTi, H1 QXi, Plus, Ultra, 16 slice H4 and HiSpeed QXi.) replaces 2269175-2. See long description for details Part #: 2269175-3 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: Longitudinal Encoder Assembly with Dynapar encoder FRU: Yes Status: Orderable Comments: Per ECR 2036924, 212008898, 12/4/2006. ** FE Assist ** If no stock or unable to order the 2269175-3 , order 2269175-2. (212001261 5/22/07) Part #: 2269175-2 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: Longitudinal Encoder Asm. NOTE:This asm. has LCC encoder/tested and sensor gap is 0.06-0.08mm FRU: Yes Status: Orderable Comments: Per ECR 2036924, 212008898, 12/4/2006
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.