Pular para o conteúdo principal
GE Healthcare · CTBrightSpeed260118119260118184

BrightSpeed HPower, Velocity or Delight HPower

Symptom

Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan

Error message

Error: 260118119 (20-0254)Heater Inverter overcurrent(inverter1) See Long Description for full error Sat Apr 15 11:57:29 2006 260118119 orp orp subsys gen exec.c 2150 Sat Apr 15 11:57:27 2006 Host : Jedi Ermes _ : 260118119 Exception Class : Soft Severity : Pri/Soft File : gen exec.c Line_ : 2150 Function : No System Function Reported Scan Type : Helical Scan: 4436/2/1 (20-0254)Heater Inverter overcurrent(inverter1)( 3times). Z6,Z7= 95h,0h

Error: 260118184 (50-0203) Heater inverter permanent overcurrent.(SW limit) See LD for error explanation Error: 260118184 (50-0203) Heater inverter permanent overcurrent.(SW limit) Code 50 Filament Heater Errors Error code: 50-0203H Message/explanation: Heater inverter permanent over current.(SW limit) Issued by the heater board when an over current has been detected and 3 restarts have been tried without success within 100 ms Potential cause: Heater board Recommended action/ Troubleshooting guide: Replace Heater board

Applies to

Models
BrightSpeed
Product
BrightSpeed HPower, Delight, Velocity or MERI HPower HP_P_G4_G_HPT Excel HP_P_G8_G_HPT Edge HP_P_G16_G_HPT Elite HP_P_G16_G_GMV Elite HP_P_P16_G_GMV Elite HP_P_M16_G_HPT BrightSpeed Method of remote software hardware identification check_config swhwinfo List of Supported Console Types : HP_P_G16_ G _HPT Note: G could be GRE, AIO or TIO - check with: cat /usr/g/config/INFO | grep CONSOLETYPE GOC4 AIO TIO List of softwares the product can run: 08BW20.1 07MW11.10 08BW44.1 ( Delight ) 09HW30.4 ( Velocity ) 09BW35.11 ( MERI - Pancake DAS ) FAQ's 1. All BrightSpeed Power Systems use the Performix tube D3186T HP_P_G4_G_HPT (Excel - H-Power Table) HP_P_G8_G_HPT (Edge -H-Power Table ) HP_P_G16_G_HPT (Elite -H-Power Table ) HP_P_G16_G_GMV (Elite -Global Table Mid Value - modified GT1700 no IMS ) Note: G could be GRE, AIO or TIO (check with: cat /usr/g/config/INFO | grep CONSOLETYPE ) BrightSpeed Elite Power Systems will only be sold as Velocity System starting in 2Q/2010. Delight Elite Power being phased out 1Q/2010 BrightSpeed Power 4 & 8 slice systems have been discontinued due to restructuring of the product lineup. Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2324118 JH4 AUXILIARY EQUIPED V2 - Used on HP and Brightspeed gantries. Not used on VCT See long description for details Part #: 2324118 Part Type: GP CONSUMABLES Part Weight: -98999.01 LBR Repairable: No Returnable: No Hazardous Material: N Description: JH4 AUXILIARY EQUIPED V2 FRU: Yes Status: Orderable Comments: FLS - JH4 AUXILIARY EQUIPED V2. TO BE USED PER HPOWER PROGRAM RSPL. MS20544 ON 2/12/03.
Area
CT