GE Healthcare · CTLightSpeedVCT260132912260122003
LightSpeed 7.x (VCT, VCT Select, Pro32)
Symptom
Customer sees "Detector Temperature Is High. Image quality may be affected". Customer sees Detector Temperature Is High. Image quality may be affected
Error message
Error: 260132912 DAS Digital Module Temperature Out of Tolerance See Long Description for more details Wed Mar 22 00:23:01 2006 Error: 260132912 Host: orp Process: orp subsys File: ThermalManager.cxx Line: 2282 Wed Mar 22 00:23:00 2006 Host : Das Ermes # : 260132912 Exception Class : Secondary Severity : Sec/Soft File : ThermalManager.cxx Line# : 2282 Function : No System Function Reported Scan Type : None/Unknown Scan: 7/2/0 DAS Digital Module Temperature Out of Tolerance Digital Module: 28 Temperature: 5003 Limits: 3800, 5000 Error: 260122003 TGP to CFC communication failure Error code: -3. See Long Description for full error Wed Mar 22 09:24:53 2006 Error: 260122003 Host: tgp Process: tgp_subsys File: gantryThermalMgr.cxx Line: 1260 Wed Mar 22 09:24:53 2006 Host : Table Gantry Processor Ermes # : 260122003 Exception Class : Secondary Severity : Pri/Soft File : gantryThermalMgr.cxx Line# : 1260 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 TGP to CFC communication failure Error code: -3.
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
- Component
- 5164727 cfc board housing assy See long description for details Part #: 5164727 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: cfc board housing assy FRU: Yes Status: Orderable Comments: Not Available
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required