GE Healthcare · CTLightSpeedVCT245043260100152179392300000692914437000
LightSpeed 7.x (VCT, VCT Select, Pro32)
Symptom
Exposure Interlock aborted scanning. Exposure Interlock aborted scanning
Error message
Error: 245043 Scanning System Aborted Scanning. See LD for complete error Tue Feb 7 07:50:12 2006 Error: 245043 Host: ctscan1 Process: scanmgr File: scanmgr_events.cxx Line: 2115 Function: System Control : Scan Control Scan Type: Axial Scan: 0/0/0 Scan Seq Id: 1000 Scanning System Aborted Scanning Error: 260100152 Exposure Interlock Opened Please refer to the long description for further details. Fri Feb 8 18:12:19 2008 Error: 260100152 Host: tgp Process: tgp_subsys File: scIsrService.c Line: 386 Fri Feb 8 18:12:19 2008 Host : Table Gantry Processor Ermes # : 260100152 Exception Class : Abort Severity : Primary File : scIsrService.c Line# : 386 Function : System Monitoring : Xray Exposure Interlocks Scan Type : Helical Scan: 17939/3/1 Exposure Interlock Opened Do Not Use - use one of the other Error: 230000069 Two, or more, unexpected events were detected during scan data transfer. The 230000069 Error has several varation see below. Sun Sep 5 11:10:31 2004 Error: 230000069 Host: darc Process: dataacq File: dip_thread.cxx Line: 396 Function: Data Acquisition : Scan Data Acquisition Scan Type: Scout Scan: 4167/1/1 Scan Seq Id: 1000 Two, or more, unexpected events were detected duri
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
- Component
- 5131930 VCT Receiver Antenna and Bracket --- Replaces (5131931) used on 7.x gantry See Long Description for picture Part # : 5131930 Part Type : GP CONSUMABLES Part Weight: -109998.90 LBR Repairable: No Returnable: No Hazardous Material: N Description: VCT Receiver Antenna and Bracket FRU: Not Available Status: Orderable 5131931 VCT Receiver. Part #: 5131931 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: VCT Receiver FRU: Yes Status: Orderable
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required