GE Healthcare · CTBrightSpeed260114039260100042
BrightSpeed HPower, Velocity or Delight HPower
Symptom
Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan
Error message
Error: 260114039 No LED MASK ACK received from Console Pushbuttons for setup of FLASHING LEDs See Long Description for details Wed Mar 11 13:12:06 2009 Error: 260114039 Host: tgp Process: tgp_subsys File: cpb_led.c Line: 77 Running: System Reset Wed Mar 11 13:12:07 2009 Host : Table Gantry Processor Ermes # : 260114039 Exception Class : Abort Severity : Primary File : cpb_led.c Line# : 77 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 No LED MASK ACK received from Console Pushbuttons for setup of FLASHING LEDs LED Setup Mask:0x1. HW/SW Error: Probable hardware, possible software, problem Error: 260100042 Cradle was NOT ready See the long description for the full error code Tue May 17 16:02:03 2005 Error: 260100042 Host: tgp Process: tgp_subsys File: scHdwRdy.c Line: 70 Tue May 17 16:02:02 2005 Host : Table Gantry Processor Ermes # : 260100042 Exception Class : Soft Severity : Sec/Soft File : scHdwRdy.c Line# : 70 Function : No System Function Reported Scan Type : Scout Scan: 106/1/1 Cradle was NOT ready
Applies to
- Models
- BrightSpeed
- Product
- BrightSpeed HPower, Delight, Velocity or MERI HPower HP_P_G4_G_HPT Excel HP_P_G8_G_HPT Edge HP_P_G16_G_HPT Elite HP_P_G16_G_GMV Elite HP_P_P16_G_GMV Elite HP_P_M16_G_HPT BrightSpeed Method of remote software hardware identification check_config swhwinfo List of Supported Console Types : HP_P_G16_ G _HPT Note: G could be GRE, AIO or TIO - check with: cat /usr/g/config/INFO | grep CONSOLETYPE GOC4 AIO TIO List of softwares the product can run: 08BW20.1 07MW11.10 08BW44.1 ( Delight ) 09HW30.4 ( Velocity ) 09BW35.11 ( MERI - Pancake DAS ) FAQ's 1. All BrightSpeed Power Systems use the Performix tube D3186T HP_P_G4_G_HPT (Excel - H-Power Table) HP_P_G8_G_HPT (Edge -H-Power Table ) HP_P_G16_G_HPT (Elite -H-Power Table ) HP_P_G16_G_GMV (Elite -Global Table Mid Value - modified GT1700 no IMS ) Note: G could be GRE, AIO or TIO (check with: cat /usr/g/config/INFO | grep CONSOLETYPE ) BrightSpeed Elite Power Systems will only be sold as Velocity System starting in 2Q/2010. Delight Elite Power being phased out 1Q/2010 BrightSpeed Power 4 & 8 slice systems have been discontinued due to restructuring of the product lineup. Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2342727-4 : 5.X Cradle encoder. This part replaces the 2342727-3 and 2342727-2 This is an absolute position encoder with no pot. See long description for details Displaying Change-Up Data: Part "2342727-2" and "2342727-3" changed to 2342727-4 Part #: 2342727-4 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: FRABA OCD H-Power String Encoder FRU: Yes Status: Orderable Comments: Added by ecr: 2020264. Please continue to use stock of 2342727-3, until depleted. -3 and -4 are mfg alts of one another. ms13452 Revised per to Asset confirmation Comments: Added per ECR2011883 to replaced the 2342727-2 The 5.x table design incorporates Absolute encoders used for cradle and elevation, instead of incremental encoders. The absolute encoder has the capability to generate a unique binary 'word' for each encoder shaft position, allowing the encoder to provide positional information instantly upon power-up, unlike incremental encoders. Therefore, potentiometers are NOT needed for cradle or elevation encoder circuits.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.