Pular para o conteúdo principal
GE Healthcare · CTLightSpeed

LightSpeed 3.X (Ultra, Plus, QX/i) - Octane

Symptom

Unable to access images in Image Works System will not scan and unable to access images in Image Works

Error message

dksc0d2vol: Device spin up failed, unable to use device -- corrective action necessary See long description /Var/adm/SYSLOG Apr 29 09:59:45 6A:ctjpm01 unix: dksc0d2vol : <6>Device not ready<6>: Not ready to perform command (asc=0x4, asq=0x0)<6> (FRU=0x2)<6> Apr 29 09:59:45 6A:ctjpm01 unix: dksc0d2vol: <6>Device spin up failed, unable to use device -- corrective action necessary

Applies to

Models
LightSpeed
Product
LightSpeed 3.X (Ultra, Plus, QX/i) - Octane Always 4 or 8 slice MDAS 308i.2_H3.1M5 LightSpeed 3.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 308I.2_H3.1M5 305.4_H3.1M4 - obsolete 303.1_H3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2169940-72UU (replaces 2169940-72) : Octane Computer 9 GB System Disk See Long Description for details Note: There are compatability problems with some consoles and disk drive firmware. Please see service note SN07112612a for details. Part #: 2169940-72UU Part Type: GP USED Part Weight: 2.43 LBR Repairable: No Returnable: No Hazardous Material: N Description: 9GB 10K RPM DISK WITH SLED FRU: Not Available Status: Orderable Comments: NOTE: MAY BE INACCURATE FOR H2 AND H3 CONSOLES. ** FE ASSIST ** 2169940-12 is no longer available can use 2169940-72 9GB or a 2169940-78 18GB Hard drives for a CTi or lightspeed QXI Octane systems as Image or OS disk. Will need to use updated LFC per 2282382-100 rev 2 or 2284291-100 rev 8. The 2169940-78 has 5000 images files where normal is 20000 if used as a CTi Image Disk. Refer to Service Note 07112612b. (212001261 8/11/09) ***OBSOLETE PART Case # 9746502*** This part has been identified as technically obsolete and no short term solution has been identified. FE must identify and order a higher or lower level spare as required. Contact your support engineering, On-line Engineering or Parts ID team should you require support. 212012173 7-24-236 Part #: 2169940-72 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 1.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: 9GB Octane 10K RPM DISK WITH SLED FRU: Yes Status: Orderable Parts Offered as: Current: 2169940-72 Used: Not Available Exchange: Not Available Comments: OCTANE 2 9GB DISK DRIVE USED ON HELIOS 2 AND 3 AND LX 9/25/03 CHECK WITH FIELD TO SEE IF THEY CAN USE THE 18GB PART# 2169940-78 IF NONE OF THE 9GB ARE AVAILIABLE
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues