GE Healthcare · CTBrightSpeed260107413
BrightSpeed HPower, Velocity or Delight HPower
Symptom
"Scan hardware Stopped Scan" with Axial Errors. "Scan hardware Stopped Scan" with Axial Errors
Error message
Error: 260107413 The Axial AC Drive reported fault code: F:13 Ground Fault See long description for full error Thu Jul 24 09:08:06 2008 Error: 260107413 Host: tgp Process: tgp_subsys File: AX2fault.c Line: 195 Thu Jul 24 09:08:06 2008 Host : Table Gantry Processor Ermes # : 260107413 Exception Class : Abort Severity : Primary File : AX2fault.c Line# : 195 Function : AXIAL CONTROL : Axial Communications Scan Type : Scout Scan: 193/1/1 %s The Axial AC Drive reported fault code: F:13 Ground Fault Check the motor and external wiring to the drive output terminals for a grounded condition.
Applies to
- Models
- BrightSpeed
- Product
- BrightSpeed HPower, Delight, Velocity or MERI HPower HP_P_G4_G_HPT Excel HP_P_G8_G_HPT Edge HP_P_G16_G_HPT Elite HP_P_G16_G_GMV Elite HP_P_P16_G_GMV Elite HP_P_M16_G_HPT BrightSpeed Method of remote software hardware identification check_config swhwinfo List of Supported Console Types : HP_P_G16_ G _HPT Note: G could be GRE, AIO or TIO - check with: cat /usr/g/config/INFO | grep CONSOLETYPE GOC4 AIO TIO List of softwares the product can run: 08BW20.1 07MW11.10 08BW44.1 ( Delight ) 09HW30.4 ( Velocity ) 09BW35.11 ( MERI - Pancake DAS ) FAQ's 1. All BrightSpeed Power Systems use the Performix tube D3186T HP_P_G4_G_HPT (Excel - H-Power Table) HP_P_G8_G_HPT (Edge -H-Power Table ) HP_P_G16_G_HPT (Elite -H-Power Table ) HP_P_G16_G_GMV (Elite -Global Table Mid Value - modified GT1700 no IMS ) Note: G could be GRE, AIO or TIO (check with: cat /usr/g/config/INFO | grep CONSOLETYPE ) BrightSpeed Elite Power Systems will only be sold as Velocity System starting in 2Q/2010. Delight Elite Power being phased out 1Q/2010 BrightSpeed Power 4 & 8 slice systems have been discontinued due to restructuring of the product lineup. Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2235342-3 AXIAL MOTOR W/ SPROCKET, HPOWER 2235342-3 AXIAL MOTOR W/ SPROCKET, HPOWER Part #: 2235342-3 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: AXIAL MOTOR W/ SPROCKET, HPOWER FRU: Yes Status: Orderable The axial motor FRU is item 1 in picture below:
- Area
- CT
Full cause & step-by-step resolution