Pular para o conteúdo principal
GE Healthcare · CTHiSpeed18333819507

HiSpeed CTi 6.X with Octane and HSA Gantry

Symptom

Scan hardware aborts scanning and allows resumes with streak artifact on images See Long Description for details and example images Scan hardware aborts scanning and allows resumes with streak artifact on images.

Error message

Error: 183338 The maximum allowable HVDC Rail voltage drop has been exceeded. See LD for complete error Tue Jul 25 09:43:35 2006 Error: 183338 Host: OBC Process: Generator File: gen_power.c 1.13 Line: 258 Function: SYSTEM MONITORING : HVDC Monitor Scan: 19507/1/1 Type: Scout Exception Level: Pri/Most Severe Time: 09:43:35:145 Log Series: 528 The maximum allowable HVDC Rail voltage drop has been exceeded. No load HVDC value: 663V 1st HVDC sample: 27V HVDC 15mS later: 27V Rotor LOV bit value: 1 From Feedback If error indicates approximately half of proper HVDC supply, this indicates strong chance of blown 700V fuse F1 or F2 on HEMRC I/F board in HEMRC

Applies to

Models
HiSpeed
Product
HiSpeed CTi 6.X HiSpeed CTi 6.X with Octane and HSA Gantry HiSpeed CTi 6.X with Octane and HSA Gantry Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2133533 CT Compact Power Distribution Unit 2133533 CT Compact Power Distribution Unit
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.