Symptom

Collimator calibration drift check has failed Cal will continue with the next step - FPA Check.

Error message

Error: 230023045 The Collimator calibration drift check has failed See the long description of the full error code Thu Aug 26 02:56:40 2004 Error: 230023045 Host: ct01 Process: dailyPrepRx File: DailyPrepUtils.c Line: 3155 The Collimator calibration drift check has failed. Full collimator calibration of all stations must be done. Cal will continue with the next step - FPA Check.

Error: 230017821 ACCS CCal: Drift check: The polynomial value has drifted by 0.470544 See the LD for the complete error code Error: 230017821 ACCS CCal: Drift check: The polynomial value has drifted Wed Aug 9 08:18:38 2006 Error: 230017821 Host: qxi1_sbc Process: accs File: collSweepProcessing.c Line: 1635 ACCS CCal: Drift check: The polynomial value has drifted by 0.470544 which is more than tolerance 0.300000

Applies to

Models
LightSpeed
Product
LightSpeed 4.X - GOC2 405L.8_H4.2M4 LightSpeed 4.X with GOC2 and MDAS 16 Gantry H4 Gantry Model # 2339985 and Global Operator Console 2 Model #'s 2366438 and 2366438-2(Hino) with HP8000 CPU (or newer) and Pegasus Recon ICE Box Product: LightSpeed16 405L.8_H4.2M4 CPU: GenuineIntel Intel(R) Xeon(TM) CPU 2.66GHz 4 2657 MHz i686 Processors Main memory size: 2048 Mbytes [root@presbyct2]# sg_map -i /dev/sg0 /dev/sda SEAGATE ST336753LW HPS2 /dev/sg1 /dev/sdb SEAGATE ST373453LW HPS1 /dev/sg2 /dev/sdc SONY SMO-F551-SD 1.05 /dev/sg3 /dev/scd0 MATSHITA DVD-RAM SW-9574S AZX8 /dev/sg4 /dev/scd1 HL-DT-ST DVD-ROM GDR8163B 0B15 [root@presbyct2]# df -h Filesystem Size Used Avail Use% Mounted on /dev/sda2 4.0G 1.7G 2.3G 41% / /dev/sda1 27M 12M 14M 44% /boot none 1008M 0 1007M 0% /dev/shm /dev/sda6 14G 3.5G 10G 25% /usr/g /dev/sda7 9.8G 753M 9.0G 8% /usr/g/sdc_image_pool /dev/sdb1 68G 28G 40G 41% /usr/g/sdc_image_pool2 Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.