GE Healthcare · CTLightSpeedVCT245207244584178570244586260001008
LightSpeed 2.X (Plus, QX/i) - GRE
Symptom
Unable to scan with a message on screen that says: Collimator/DAS reset failed.
Error message
Error: 245207 Software has detected an invalid or mistached DAS type. This will prevent the system from scanning. see long description Mon Apr 3 08:29:12 2006 Error: 245207 ls16 scanmgr validate request.cxx 378 Software has detected an invalid or mistached DAS type. This will prevent the system from scanning. Error: 244584 Collimator/DAS reset failed See Long Description for complete error Error: 244584 Host: qx-ct Process: fwmgr File: HwManager.c Line: 1536 Collimator/DAS reset failed. Error: 178570 Coll/Das controller reports failure in initializing tasks See Long Description for complete error Mon Mar 28 03:39:48 2005 Error: 178570 Host: ct04 Process: fwmgr File: HwSetUp.c Line: 3178 Function: No system function reported Scan Type: Not scanning Scan: 0/0/0 Scan Seq Id: 0 OBC_SS timeout: Coll/Das controller reports failure in initializing tasks. Error: 244586 the following hardware subsystems reported a failure when initializing: OBC SS (Coll/Das) See the Long Description for the complete error code Error: 244586 Host: qx-ct Process: fwmgr File: HwSetUp.c Line: 2235 Function: No system function reported Scan Type: Not scanning Scan: 0/0/0 Scan Seq Id: 0 The following hardware
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 2.X (Plus, QX/i) - GRE always an SDAS 4 slice H2_P_S4 (SDAS4 and H2 gantry) Lightspeed 2.X - GRE Method of visual identification Small rounded gantry similar in appearance to LS 3.X and 4.X. Always has an SDAS 4 slice Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console Types GOC3 Will have a DARC, SDDA and IG GOC4 Will have a DARC 2 and IG List of softwares the product can run 07MW11.10 SP3 06MW29.7 - obsolete 06MW03.5 - obsolete 05MW14.5 - obsolete 04MW44.9 - obsolete 04MW13.7 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2225218 DCB HOUSING ASM for SDAS See LD for details Part #: 2225218 Part Type: GP REPAIRABLES Part Weight: 0.66 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: DCB HOUSING ASM (SDAS) FRU: Yes Status: Orderable ====================== Parts Offered as: Current: 2225218 Harvested: Not Available Exchange: Not Available Comments: FLS - DCB ASM, P/N 2225218 7/31/02- C2R COMPLETED PER M DIGGINS, CHANGE ORDER TO T3202KJ BEFORE SHIPPING. MS12464 CAN'T ORDER DAS ASSY SEE T3202KJ 1/2/02 VEN HAS 20 PC MINIMUM BUY PER PO AND 60 DAY LEAD TIME. MS09121 *** CD & S NOTE **** BEFORE REPLACING THE SDAS DAS CONTROL BOARD (DCB), MEASURE THE +5V ON THE BOARD FROM TP-1 - TP-2. VERIFY THIS IS ABOVE4.8 VOLTS. VOLTAGE LOWER THANTHIS MAY CAUSE INTERMITTENT OPERATIONS. IF THE VOLTAGE ISOK ON THE CONVERTER CARD CHASSCHASSIS, YOU MAY HAVE A CABLE PROBLEM. MS30863 5/31/02 GELS - CSO0111459 FRANCE F-45071720002 /04/01/TS FIBER OPTIC FREQ 50 MHZ / NEEDS DIP BOARD 2275604 OR 2245261 PDB 02/01/21,DCB HOUSING ASM
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.