BrightSpeed Select or Delight Select

Symptom

Alignment lights do not turn on, nor does Gantry rotate. Errors in system log. Alignment lights don't come on, nor does Gantry rotate. Errors in system log.

Error message

Error: 260107221 Gantry (Axial) LOOP CONTACTOR state is Still OPEN. see Long Description for details Wed Apr 9 21:12:15 2008 Error: 260107221 Host: tgp Process: tgp_subsys File: amc_mgr.c Line: 3318 Wed Apr 9 21:12:15 2008 Host : Table Gantry Processor Ermes # : 260107221 Exception Class : Abort Severity : Primary File : amc_mgr.c Line# : 3318 Function : AXIAL CONTROL : Axial Control Scan Type : None/Unknown Scan: 0/0/0 Gantry (Axial) LOOP CONTACTOR state is Still OPEN. The Loop Contactor status is NOT changing to the CLOSED state. (AMC State:9 Cmd:21).

Applies to

Models
BrightSpeed
Product
BrightSpeed Select, Delight Select or MERI Select Elite-16, Edge-8, Excel-4 HP_S_G16_G_HLT HP_S_G8_G_HLT HP_S_G4_G_HLT BrightSpeed Select or Delight Select (Hivalue Lite) 4/8/16 Slice Scanner BrightSpeed Delight Select Systems started to ship in first Quarter or 2009 with AIO (All in One) Console. AIO console uses HP8400 and NO DARC or IG's. Recon time is 6 FPS. If customer orders Fluoro option system requires one IG. The 16 FPS Upgrade in Not Available for Select systems. BrightSpeed Meri Select Systems ( Pancake Das ) started to ship in 3rd quarter of 2010, 4/8/16 Slice Scanner HiValue 16 Lite −−> BrightSpeed Elite Select HiValue 8 Lite−−> BrightSpeed Edge Select HiValue 4 Lite−−> BrightSpeed Excel Select BrightSpeed Select Scope: - 3.5MHu Solarix tube - JEDI 60 dc Lite - Modified NP Table Key Features: - 16x0.625mm(16x1.25) for Lite 16 - 8x1.25/8x2.5mm for Lite 4/8 - Rotation Speed: 0.8 s - Gantry: 2040mm width, 70cm Bore - Table: 1337@ISO-140mm scannable range - Recon: 3fps (w/o IG @ 2G Memory(4/8), 4G Memory(16)), optional: 6fps(w/ 1IG ) - Power: 42KW JEDI power See the attached Service Intro Planner for details Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2351490 18-32A CONTACTOR / RELAY See long description for details Part #: 2351490 Part Type: GP CONSUMABLES Part Weight: 1.32 LBR Repairable: No Returnable: No Hazardous Material: N Description: 18-32A CONTACTOR FRU: Yes Status: Orderable Comments: FLS - 18-32A CONTACTOR. ADDED PER NGPDU RSPL FROM ASIA. MS20544 ON 1/2/03.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues