GE Healthcare · CTLightSpeed200002632
LightSpeed 4.X - GRE
Symptom
OPTION STATUS LOG box popped up on screen "All FlexTrial Options are uninstalled and the user can no longer use these options" See Long Description for Details and pop up attention box. Option status log box popped up on screen "All FlexTrial Options are uninstalled and the user can no longer use these options" with Error: 200002632 System Back dated. SW Options will be disabled.
Error message
Error: 200002632 System Back dated. SW Options will be disabled. See Long Description for Details. Mon Dec 31 18:03:15 2001 Error: 200002632 Host: darc Process: dataacq File: swokutl.cxx Line: 784 *** Error: System Back dated. SW Options will be disabled. Please change the system date to current date or contact OLC
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2334572 The 3 VOLT BATTERY 1" DIAMETER for CPU's aka CR2032 See long description for details 2334572 The 3 Volt Battery, 1" in diameter, for CPU's also known as CR2032 This may be purchased locally at a variety of locations, including Radio Shack, Walgreens, etc. Part #: 2334572 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: 3 VOLT BATTERY 1" DIAMETER FRU: Yes Status: Orderable Comments: FLS - NEWARK # 13C1019 # MANUF. CPI CANADA VC#726749 PART# 742290100 3 VOLT BATTERY 1" DIAMETER U/O LORAD M4 MICROPROCESSOR MS29328 - LMG - 2/18/02
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.