GE Healthcare · CTHiSpeed186271
Difficulty in positioning the cradle.
Symptom
Difficulty in positioning the cradle. Cradle will not drive normally and scan aborts with errors.
Error message
Error:186271 Code: 12-1001-17 Error detected by:Cradle acceleration/deceleration time out. Or total cradle movement time out on Helical Scan. See long description for full error Fri Apr 10 21:20:21 2009 Error:186271Host: IRIS Proc: tgp_out tgp_errors.c 296Helical Scan Error Code : 12-1001-17 Error detected by : TGP(TP) Cradle acceleration/deceleration time out. Or total cradle movement time out on Helical Scan.
Applies to
- Models
- HiSpeed
- Product
- HiSpeed YMS Series HiSpeed Series DXi, FXi, LXi, ZXI, and NXI Including Single / Twin detector and Glass / Metal Tubes. See Long Description for Details Subsystem/Components DXI (NP) FXI,LXI (NP+) ZXI (NP++ NXI (NP+ Twin NXI (NP++ Twin OC O2 running Unix or upgraded to Linux PC Gantry Alighnment Lights Halogen Laser Laser Laser Laser Mechanics Common Electronics firmware is different twin Det/Das Twin Det/DAS Table IMS no Yes Yes Yes Yes Generator Jedi Jedi NP++ Jedi Jedi NP++ Jedi Tube D3142T D3142T D3152T D3142T D3152T PDU Common Options List for V/R 5.5, 6, 7, 9, and 10 System 3D Power 440ma Max Image Disk 20000 Smart Prep Fast Scam 0.7 sec Performix 530 cooling Smart Recon Fast Scan 0.8 sec Asymmetric Scan SmartView Fast Scan 1.0 sec Cardiac gating DemtaScan Helical 30 Max Smart Addition Navigator Helical 60 Max Respiratory Gating Helical Plus Helical 120 Max Reformat Cpmmect {rp Raw Data 1000 max VolumeRendering2 ImageDosk Remote Tilt CT Colono Pioneer Mod Enhanced Cooling The following Options name changes Option (V5.5x,V6.0x,V7.x,V9.x,V10x) Option (V8.1x and V11.x) 3D >> Volume Analysis2 Navigator >> Navigator2,VolumeRendering2,VolumeAnalysis2 CtPerfusion >> CTPersusion2 Reformat >> Functool Options V8.1x and V11.x VolumeAnalysis2 Power 440ma ImageDisk 2000 Helical Plus Smart Prep Fast Scan 0.7 sec Performix 530 cooling Connect Pro Smart Recon Fast Scan 0.8 sec Asymmetric Scan ImageDisk SmartView Fast Scan 1.0 sec Cardiac Gating Pioneer MOD* DentaScan Helical 30 Smart Addition Power 200 ma max Navigator2/VolumeRendering2/ VolumeAnalysis Helical 60 Respiratory Gating Power 250 ma Max Power 300 ma Max Helical 120 Thickness 0.5mm CTColon Power 350 ma Max Raw data 1000 CtPerfusion2 AdvVesselAnalysis Remote Tilt Extended FOV Functool Enhance Cooling Performix Tube VolumeRendering2
- Component
- 2201876 STEP MOTOR ASSY - used on YMS Xi series tables See Long Description Part #: 2201876 Part Type: GP CONSUMABLES Part Weight: 10.01 LBR Repairable: No Returnable: No Hazardous Material: N Description: P9235SA STP MOTOR ASSY FRU: Yes Status: Orderable Comments: FLS - STEP MOTOR ASM U/O CT HISPEED-YMS TABLE M3 FW20 '98 MS1101 3/98 ***GPRS DISCOUNTABLE PART 9/08*** 212008582 07-OCT-2008 GELS - KANBAN CSS F00,STEP MOTOR ASSY 2201798 CT Hispeed YMS cradle belt See long description for details Part #: 2201798 Part Type: GP CONSUMABLES Part Weight: 0.99 LBR Repairable: No Returnable: No Hazardous Material: N Description: BELT FRU: Yes Status: Orderable Comments: FLS - BELT U/O CT HISPEED-YMS TABLE M3FW20 '98 MS1101 3/98 2201854 Cradle Roller Belt see long descriptions for details Part #: 2201854 Part Type: GP CONSUMABLES Part Weight: 0.99 LBR Repairable: No Returnable: No Hazardous Material: N Description: BELT FRU: Yes Status: Orderable Comments: FLS - BELT U/O CT HISPEED-YMS TABLE M3 FW20 '98 MS1101 3/98 GELS - KANBAN CSS F00,BELT
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required