Symptom
Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan
Error message
Error: 260115244 MTCB cradle moving direction error See the Long Description for the complete error code Thu Dec 28 10:59:18 2006 Error: 260115244 Host: tgp Process: tgp subsys File: mtcb cradle monitor.c Line: 823 Thu Dec 28 10:59:18 2006 Host : Table Sub-System Ermes # : 260115244 Exception Class : Secondary Severity : Pri/Soft File : mtcb cradle monitor.c Line# : 823 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 MTCB cradle moving direction error. Possible Encoder, cabling or MTCB board issue. Current Speed = -7300 Target Speed = 75000 Target Position = 1703000 Current Position = 101663
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 5.X (80/100kW) LightSpeed 5.X (80/100kW) ~ Use with both MDAS and GDAS ~ GRE Console HP_H_G16 HP_H_M16 HP_O_M16 LightSpeed 5.X (100kW) GDAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2342727-4 : 5.X Cradle encoder. This part replaces the 2342727-3 and 2342727-2 This is an absolute position encoder with no pot. See long description for details Displaying Change-Up Data: Part "2342727-2" and "2342727-3" changed to 2342727-4 Part #: 2342727-4 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: FRABA OCD H-Power String Encoder FRU: Yes Status: Orderable Comments: Added by ecr: 2020264. Please continue to use stock of 2342727-3, until depleted. -3 and -4 are mfg alts of one another. ms13452 Revised per to Asset confirmation Comments: Added per ECR2011883 to replaced the 2342727-2 The 5.x table design incorporates Absolute encoders used for cradle and elevation, instead of incremental encoders. The absolute encoder has the capability to generate a unique binary 'word' for each encoder shaft position, allowing the encoder to provide positional information instantly upon power-up, unlike incremental encoders. Therefore, potentiometers are NOT needed for cradle or elevation encoder circuits.
- Area
- CT
Full cause & step-by-step resolution