Symptom

"Scan hardware Stopped Scan" with collimator errors. "Scan hardware Stopped Scan" with collimator errors

Error message

Error: 260003118 CAM A drive tracking detected the current motor position is out of the tolerance band for a successfull move See Long Description for complete error. Error: 260003118 CAM A drive tracking detected the current motor position is out of the tolerance band for a successfull move Mon Aug 6 09:44:18 2001 Error: 260003118 Host: HSC Process: CCB Firmware File: ccb_drive_timers.c Line: 560 Running diagnostic: Collimator Functional Function: XRAY GENERATION : Xray Collimation and Filtration Scan: 50172/4/0 Type: None/Unknown Exception Level: Sec/Soft Time: 09:44:18:540 Log Series: 90 CAM A drive tracking detected the current motor position is out of the tolerance band for a successfull move Start Position = 3995um 165cnt Desired Position = 2um -863cnt Current Motor Position = 4336um -119cnt Current Encoder Position= 4413um -67cnt Retry Tolerance = 74um Wed Jan 16 06:38:30 2008 Error: 260003118 Host: HSC Process: CCB Firmw are File: ccb_drive_timers.c Line: 560 Function: XRAY GENERATION : Xray Collimation and Filtration Scan: 50066/8/46 Type: Axial Exception Level: Sec/Soft Time: 06:38:30:435 Log Series: 543 Filter drive tracking detected the current motor position is out of 

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 2.X (Plus, QX/i) - GRE always an SDAS 4 slice H2_P_S4 (SDAS4 and H2 gantry) Lightspeed 2.X - GRE Method of visual identification Small rounded gantry similar in appearance to LS 3.X and 4.X. Always has an SDAS 4 slice Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console Types GOC3 Will have a DARC, SDDA and IG GOC4 Will have a DARC 2 and IG List of softwares the product can run 07MW11.10 SP3 06MW29.7 - obsolete 06MW03.5 - obsolete 05MW14.5 - obsolete 04MW44.9 - obsolete 04MW13.7 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
5422592 (replaces 2125453-2) Coupling and Motor Assembly, RoHS Helios Collimator - used on LS 2X - 5X collimators see long description for details Part #: 5422592 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: Coupling and Motor Assembly, RoHS Helios Collimator FRU: Yes Status: Orderable Old FRU Compatible No Comments Added per to ECR2118136 by sso#305024364 Aug 30, 2011 Additional FRU Needed No Old FRU Scrapped No Old FRU Upgradable No Use on Hands No Use Demand History Yes Ultimate Prime 5422592 New FRU Backward Compatible Yes Old FRU Usable Yes Change Type Direct ECO Number ECR2118136 Part #: 2125453-2 Description: HELIOS COLLIMATOR, CAM MOTOR ONLY FRU: Yes Status: Orderable These motors are used in collimators for LS 1X - 5X
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.