Pular para o conteúdo principal
GE Healthcare · Nuclear MedicineInfinia

Unable to perform Greset.

Symptom

Unable to perform Greset. Gantry reboots properly CIB fails diagnostics Unable to perform Greset. Gantry reboots properly. CIB fails diagnostics

Error message

RFR_HW_NOT_FOUND, ERROR 3066 RFR HW NOT FOUND:.:No HW is found. Cant work in real mode:.:None:.: No HW is found. Cant work in real mode:.:Trace:{0) TLT 3066@ERROR MSG

RFR_EVENT_INDEX_MISMATCH, ERROR 3074 RFR EVENT INDEX MISMATCH:.:Communication/Synchronization error with the Camera Interface Board.:.:1. Perform full shutdown. 2. Perform PC reboot.:.: Event entry and event Z index are mismatched. Event entry is Undefined event Z index is Undefined:.:Trace:{2} TLT 3074@ERROR MSG

Applies to

Models
Infinia
Product
Infinia HE4 Infinia II Hawkeye 4 & Infinia II GP (Infinia running Windows XP software)
Component
5177387 Camera Interface Board (CIB) 5177387 Camera Interface Board (CIB)
Area
Nuclear Medicine