GE Healthcare · CTLightSpeed2600073915010232719
LightSpeed 4.X - Octane
Symptom
"Scan hardware Stopped Scan" with DAS power errors. "Scan hardware Stopped Scan" with DAS power errors
Error message
Error: 260007391 The Data Acquisition System Has Detected An Error In The +5V Analog DAS Power Supply. Same error for -5v Analog DAS Power Supply. Error: 260007391 The Data Acquisition System Has Detected An Error In The +5V Analog DAS Power Supply. Same error # for -5v Analog DAS Power Supply. Error: 260007391 The Data Acquisition System Has Detected An Error In The +5V Analog DAS Power Supply Wed Feb 22 08:50:01 2006 Host : Das Ermes # : 260007391 Exception Class : Abort Severity : Secondary File : pwrtst.c Line# : 90 Function : No System Function Reported Scan Type : None/Unknown Scan: 50102/3/0 The Data Acquisition System Has Detected An Error In The +5V Analog DAS Power Supply . Measured: 1590 mV, Expected range: 4750 to 5250 mV Tue Aug 22 15:51:09 2006 260007391 4 orp orp_subsys pwrtst.c 90 Tue Aug 22 15:51:08 2006 Host : Das Ermes # : 260007391 Exception Class : Abort Severity : Secondary File : pwrtst.c Line# : 90 Function : No System Function Reported Scan Type : None/Unknown Scan: 32719/3/0 The Data Acquisition System Has Detected An Error In The -5V Analog DAS Power Supply . Measured: -4748 mV, Expected range: -4751 to -5249 mV
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 4.X - Octane sw_version: 405I.2_H4.0M5 LightSpeed 4.X - Octane Gantry Model # 2339985 Console Model #'s 2341923 and 2341923-2(D3D) check_config shows LightSpeed 16 / LightSpeed 16 Product: LightSpeed16 sw_version: 405I.2_H4.0M5 or newer machine_type: Octane host_id: 3232297858 tube_type: 15 Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2218570 BOX OF FUSES for HSA, CTI, and LS 1X - 4X Systems See long description for details 2218570 BOX OF FUSES for HSA, CTI, and LS 1x - 4x Systems Fuse KIT
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.