GE Healthcare · CTLightSpeedVCT290010002290010005290009015290009030
LightSpeed 7.x (VCT, VCT Select, Pro32)
Symptom
Reconstruction Error detected during boot-up - reconSelftest. Reconstruction Error detected during boot-up - reconSelftest . Pop ping pop up box comes up. Attention: Reconstruction Process has stopped.
Error message
Error: 290010002 DR-IG communication error of writing data to socket . Error: 290010002 DR-IG communication error of writing data to socket Error: 290010005 Socket error: socket is terminated because IG dies (error text varies) See Long Description for error examples. Tue Mar 16 10:40:09 2004 290010005 darc data_restore request_object.cxx 503 Socket error: socket is terminated because IG dies Fri Nov 16 19:22:57 2007 Error: 290010005 Host: darc Process: data_restore File: request_object.cxx Line: 556 Socket error: socket event from IG Error: 290009015 Recon connect timer triggered on 3 ( could also be 1 or 2 ) See LD for full error Error: 290009015 Recon connect timer triggered on 3 Tue Aug 22 07:26:18 2006 Error: 290009015 Host: darc Process: recon_control File: recon_control_cb.cxx Line: 1605 Recon connect timer triggered on X X = the number of ig that failed ( i.e. 1 , 2 or 3) Error: 290009030 Failed to stop image_generation process See long description for full error Mon Nov 19 06:31:41 2007 Error: 290009030 Host: darc Process: recon_control File: IGControl.cpp Line: 591 Failed to stop image_generation process.
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required