GE Healthcare · CTLightSpeed230017812000500230017801
LightSpeed 3.X (Ultra, Plus, QX/i) - GOC2
Symptom
Collimator calibration drift check has failed Cal will continue with the next step - FPA Check.
Error message
Error: 230017812 ACCS CCal: Error: too much variation in ratio data. Limit 0.000500 See Long Desciption for complete error Mon May 30 05:41:49 2005 Error: 230017812 Host: cxi1 Process: accs File: collSweepProcessing.c Line: 491 ACCS CCal: Error: too much variation in ratio data. Limit 0.000500 Error: 230017801 ACCS CCal: Sweep Processing failed. Status = -1 See the Long Description for the full error code Tue Nov 30 02:33:04 2004 Error: 230017801 Host: k2_ct2_ls Process: accs File: ccal_prep.c Line: 661 ACCS CCal: Sweep Processing failed. Status = -1
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 3.X (Ultra, Plus, QX/i) - GOC2 This is HP CPU host running LINUX with Pegasus IG. 308L.8_H3.3M4 ~ 4 or 8 slice MDAS. ~ Could be an Ultra, Plus or QXi system. Must run swokinstall -p to verify. ~ See Long description for details . LightSpeed 3.X MDAS4 Gantry and GOC2 Gantry - 2281177-2, DAS - 2272768, Console – 2341104-2, Table - 2276776, PDU - 2269902 Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Area
- CT
Full cause & step-by-step resolution