Pular para o conteúdo principal
GE Healthcare · CTLightSpeed18310918763

LightSpeed 3.X (Ultra, Plus, QX/i) - Octane

Symptom

Anode or Cathode overcurrent errors when making X-ray. Anode overcurrent errors during tube warm up. Getting Diagnostic Errors 214321 and 214253 when running KV loop Diagnostic Thu May 18 11:38:42 1995 Suite: SAIC Host: OBC Proc: Diag Exec Error: 214321 File: technic_monitor.c 1.36 Method: technic_monitor Line: 1205 Function: Diagnostics XRAY GENERATION : MA Loop Test Name: Manual KV Unique ID: 48133 Iteration: 1 Exc. Level: Hard Time: 73779 Error Type: Hardware The maximum mA limit has been exceeded. Expected mA: 0mA Actual mA: Anode=121mA Cathode=121mA Maximum mA allowed: 5mA Thu May 18 11:35:28 1995 Suite: SAIC Host: OBC Proc: Diag Exec Error: 214253 File: isr1_handler.c 1.28 Method: obc_isr1_handler Line: 271 Function: Diagnostics XRAY GENERATION : KV Loop Test Name: XRay - Manual Unique ID: 48030 Iteration: 1 Exc. Level: Hard Time: 35142 Error Type: Hardware KV fault detected: Anode overcurrent. KV address: FFCFF3H Bit: D4

Error message

Error: 183109 KV fault detected: Anode overcurrent See Long Description for complete error Tue Nov 30 15:38:17 2004 Error: 183109 Suite: CT01 Host: OBC Process: Generator File: gen_abort.c 1.33 Method: No Method Line: 248 Function: XRAY GENERATION : KV Loop Scan: 18763/2/1 Type: Helical Exception Level: Pri/Most Severe Time: 15:38:16:120 Log Series: 991 KV fault detected: Anode overcurrent. Fault latch address = ffcff3 hex, Bit = D4

Applies to

Models
LightSpeed
Product
LightSpeed 3.X (Ultra, Plus, QX/i) - Octane Always 4 or 8 slice MDAS 308i.2_H3.1M5 LightSpeed 3.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 308I.2_H3.1M5 305.4_H3.1M4 - obsolete 303.1_H3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
46-195120G21 HV CABLES 3 PIN 7 FT See LD for details Part #: 46-195120G21 Part Type: GP CONSUMABLES Part Weight: 3.19 LBR Repairable: No Returnable: No Hazardous Material: N Description: HV CABLES 3 PIN 7 FT FRU: Yes Status: Orderable
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues