GE Healthcare · CTLightSpeed26011631650119244634260116313244540244520
LightSpeed RT16 & RT16 Xtra with GRE and GDAS16 gantry.
Symptom
Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan
Error message
Error: 260116316 Communications are DOWN between subsystems See LD for full error Mon Jun 13 14:34:09 2005 Error: 260116316 Host: tgp Process: tgp_subsys File: lan_watchdog.c Line: 904 Mon Jun 13 14:34:03 2005 Host : Table Gantry Processor Ermes # : 260116316 Exception Class : Safe State Severity : Primary File : lan_watchdog.c Line# : 904 Function : No System Function Reported Scan Type : None/Unknown Scan: 7754/5/0 Communications are DOWN between subsystems Or Sun Dec 17 08:26:45 2006 Error: 260116316 Host: tgp Process: tgp_subsys File: lan_watchdog.c Line: 996 Sun Dec 17 08:26:39 2006 Host : Table Gantry Processor Ermes # : 260116316 Exception Class : Safe State Severity : Secondary File : lan_watchdog.c Line# : 996 Function : No System Function Reported Scan Type : None/Unknown Scan: 50119/1/0 Slip-Ring Communications are DOWN between subsystems Error: 244634 Subsystem State; DCB:up, CCB:up, TCB:up, JEDI:up, - TGP/ORP:down - , DEVICES:up GOC6 version TGP/ORP:DOWN See Long Description Wed Feb 17 23:30:55 2010 Error: 244634 Host: bay46 Process: fwmgr File: fwmgr_utils.c Line: 677 Function: No system function reported Scan Type: Not scanning Scan: 0/0/0 Scan Seq Id: 0 Subsystem S
Applies to
- Models
- LightSpeed
- Product
- LightSpeed RT16 & RT16 Xtra and Vision WB_H_G16_G_GTL = RT16, Xtra WB_H_G16_G_GTL650 = Vision WB_H_G16_G_GTM = RT16, Xtra or Vision LightSpeed RT16 & RT16 Xtra with GRE and GDAS16 gantry Links to the Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Software
- 08BW44.1 Software for all BrightSpeed Systems with AIO Console and all RT16 Systems with GOC4 Console See Long Description 08BW44.1 Software is used on all BrightSpeed Systems with AIO Consoles and all RT16 Systems with GOC4 Consoles. Software collector called “08BW44.X” p/n 5328867 contains the following items: 5323837 - LightSpeed Xtream Console Operating System Version - GEMS Linux 6.3.8 5323838 - LightSpeed Applications Software DVD Version 08BW44.X 5328828 - Xtream Serial-Over-LAN Service Software Version - 2.0 Supports: Westville or Jarrell DARC/VDARC 5323839 - Class C 5323840 - Class M NOTE: Class C/M install CD are not shipped with system. NOTE: The Xtream Serial-Over-Lan Service Software is not used during the LFC unless instructed via a pop-up message box "SOL Service - ...run the Serial Over LAN Service CD...". This CD is loaded directly into the DARC Node. Then press the reset (or cycle power) at the front panel of the DARC Node as required. This CD will auto-eject when the SOL load process is complete. NOTE: The Xtream Serial-Over-Lan CD is not used for All-In-One Console.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required