Symptom

Attention: The Scanning Hardware is not available. Please see long description. Attention: The Scanning Hardware is not available, please wait for the hardware reset successfull status posted in the feature status area on the right monitor. If the hardware reset fails, then reset the scan hardware from the Service Desktop System Resets button.

Error message

Error: 244634 Subsystem State; DCB:up, CCB:up, TCB:up, JEDI:up, - TGP/ORP:down - , DEVICES:up GOC6 version TGP/ORP:DOWN See Long Description Wed Feb 17 23:30:55 2010 Error: 244634 Host: bay46 Process: fwmgr File: fwmgr_utils.c Line: 677 Function: No system function reported Scan Type: Not scanning Scan: 0/0/0 Scan Seq Id: 0 Subsystem State; DCB:up CCB:up TCB:up JEDI:up TGP/ORP:down DEVICES:up

Error: 244540 Scan Seq Id: 0 Immediate Request cannot be honored since the firmware is in the DOWN state See Long Description for complete error Sun Sep 26 08:36:46 2004 Error: 244540 4 000ctge04 fwmgr FwMgrEventHandler.c 1517 Running diagnostic: Scan Data Path Functional Function: No system function reported Scan Type: Not scanning Scan: 0/0/0 Scan Seq Id: 0 Immediate Request cannot be honored since the firmware is in the DOWN state

Applies to

Models
LightSpeed, VCT
Product
LightSpeed RT16 & RT16 Xtra and Vision WB_H_G16_G_GTL = RT16, Xtra WB_H_G16_G_GTL650 = Vision WB_H_G16_G_GTM = RT16, Xtra or Vision LightSpeed RT16 & RT16 Xtra with GRE and GDAS16 gantry Links to the Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.