GE Healthcare · CTLightSpeedVCT260007768
LightSpeed 4.X - Octane
Symptom
Can't scan with Pink Warning: Software has detected an invalid DAS configuration.
Error message
Error: 260007768 DAS Converter Boards are not installed in valid configuration See LD for complete error Mon Jan 23 07:02:45 2006 Error: 260007768 Host: DAS Process: DCB Firmware File: cnvmgr.c Line: 197 Function: DATA ACQUISITION : Scan Data Acquisition Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Soft Time: 07:02:445:750 Log Series: 30 DAS Converter Boards are not installed in valid configuration. The Following Card List Indicates Potential Missing Boards
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 4.X - Octane sw_version: 405I.2_H4.0M5 LightSpeed 4.X - Octane Gantry Model # 2339985 Console Model #'s 2341923 and 2341923-2(D3D) check_config shows LightSpeed 16 / LightSpeed 16 Product: LightSpeed16 sw_version: 405I.2_H4.0M5 or newer machine_type: Octane host_id: 3232297858 tube_type: 15 Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required