LightSpeed 7.x (VCT, VCT Select, Pro32)
Symptom
Scan hardware stopped Scan A Digital Module Has Exceeded The Maximum Temperature See Long Description: Scan hardware stopped Scan Error: 260132744 A Digital Module Has Exceeded The Maximum Temperature and Error: 260132802 The Data Acquisition System Has Detected An Over Temperature Condition. and Error: 260122003 TGP to CFC communication failure Error code: -3. Thu Mar 23 08:09:33 2006 Error: 260122003 Host: tgp Process: tgp_subsys File: gantryThermalMgr.cxx Line: 761 Thu Mar 23 08:09:33 2006 Host : Table Gantry Processor Ermes # : 260122003 Exception Class : Secondary Severity : Pri/Soft File : gantryThermalMgr.cxx Line# : 761 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 TGP to CFC communication failure Error code: -3. ============================================================ Wed Mar 22 01:26:44 2006 Error: 260132744 orp orp_subsys Ifb.cxx 501 Wed Mar 22 01:26:43 2006 Host : Das Ermes # : 260132744 Exception Class : Secondary Severity : Sec/Soft File : Ifb.cxx Line# : 501 Function : No System Function Reported Scan Type : None/Unknown Scan: 11/2/0 Digital Interface Board(s) Have Detected An Digital Module Has Exceeded The Maximum Temperature DIFB : 28 Digital Modules: 52-56 Temperature : 6506 =============================================================== Wed Mar 22 01:26:44 2006 Error: 260132802 orp orp_subsys IfbManager.cxx 1744 Wed Mar 22 01:26:43 2006 Host : Das Ermes # : 260132802 Exception Class : Abort Severity : Primary File : IfbManager.cxx Line# : 1744 Function : No System Function Reported Scan Type : None/Unknown Scan: 11/2/0 The Data Acquisition System Has Detected An Over Temperature Condition. All Digital Module Power Has Been Disabled Until A Hardware Reset Is Performed Wed Mar 22 00:23:01 2006 Error: 260132912 Host: orp Process: orp subsys File: ThermalManager.cxx Line: 2282 Wed Mar 22 00:23:00 2006 Host : Das Ermes # : 260132912 Exception Class : Secondary Severity : Sec/Soft File : ThermalManager.cxx Line# : 2282 Function : No System Function Reported Scan Type : None/Unknown Scan: 7/2/0 DAS Digital Module Temperature Out of Tolerance Digital Module: 28 Temperature: 5003 Limits: 3800, 5000
Error message
Error: 260132912 DAS Digital Module Temperature Out of Tolerance See Long Description for more details Wed Mar 22 00:23:01 2006 Error: 260132912 Host: orp Process: orp subsys File: ThermalManager.cxx Line: 2282 Wed Mar 22 00:23:00 2006 Host : Das Ermes # : 260132912 Exception Class : Secondary Severity : Sec/Soft File : ThermalManager.cxx Line# : 2282 Function : No System Function Reported Scan Type : None/Unknown Scan: 7/2/0 DAS Digital Module Temperature Out of Tolerance Digital Module: 28 Temperature: 5003 Limits: 3800, 5000
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
- Component
- 5272157-2 - Cherokee 48Vdc 1KW supply with Ferrite filter - used on 7.X Gantry Stationary frame - replaces 5125186 and 5272157 See Long Description for details Part #: 5272157-2 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: Cherokee 48Vdc 1KW supply with Ferrite filter FRU: Yes Status: Orderable Comments: Per ECR 2086962, 212008898, 7/9/09. Part #: 5272157 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: Cherokee 48Vdc 1KW supply with modified input filtering. FRU: Yes Status: Orderable Comments: Per ECR 204282, 212008898, 12/4/07 Part #: 5125186 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: PM3326A-6-1-2-C Pioneer Magnetics 48V 1kW Power Supply FRU: Yes Status: Orderable Comments: Added per VCT RSPL, ms13452 041209 48 volts DC power supply used on VCT Stationary frame. If Rotating frame 48 volts DC supply is needed, order 5126259. (added by 212001261 9/12/05)
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required